- Purifier HEPA Filtered Enclosures & Purifier Class I Filtered Enclosures User's Manual

11
Cha
p
ter 3: Gettin
g
Starte
d
0.36" of the front of the work surface. Mounting holes are
provided in the Labconco accessory work surfaces to secure the
enclosure.
Work Surface Specifications
The work surface should be smooth, rigid, and durable, such as a
chemical-resistant epoxy resin. The surface should be non-porous
and resistant to the biohazards, powders, solvents and chemicals
used in conjunction with the Purifier Filtered Enclosure. The work
surface should also contain a dished recessed area for containing
primary spills.
Work Surface and HEPA Filtered Enclosure Installation
1. Level the base cabinets and the work surface. Work
surface should be placed flush with the front of the base
cabinet as shown in Figure 3-1.
2. Position the work surface in its intended location and with
the front of the work surface towards you. (Rear mounting
holes are located close to the rear edge.)
3. Secure the work surface to the base cabinet with a
structural adhesive or silicone sealant.
4. Insert the supplied mounting screws in the four holes.
Allow a minimum of 1/8" clearance under the head of the
screw for positioning the enclosure.
5. Place the enclosure on the work surface and slide the rear
flange and front air foil flanges under the mounting screw
heads.
6. Tighten the four screws to complete the installation.