Product Manual
Table Of Contents
- 1 Safety Directions
- 2 Description of the System
- 3 User Interface
- 4 Operation
- 4.1 Main Menu
- 4.2 System Information
- 4.3 Setting Up the TPS Instrument
- 4.4 Setting up for Remote Control (with the RadioHandle)
- 4.5 Connecting to a Personal Computer
- 4.6 Power Functions
- 4.7 Batteries
- 4.8 Working with the Memory Device
- 4.9 Working with Bluetooth
- 4.10 LED Indicators
- 4.11 Guidelines for Correct Results
- 5 Settings
- 6 Apps
- 7 Care and Transport
- 8 Technical Data
- 8.1 Angle Measurement
- 8.2 Distance Measurement with Reflectors
- 8.3 Distance Measurement without Reflectors
- 8.4 Distance Measurement - Long Range (LO mode)
- 8.5 Automatic Prism Aiming AiM
- 8.6 Prism Search (Scout) - only available at Zoom90
- 8.7 Conformity to National Regulations
- 8.8 General Technical Data of the Product
- 8.9 Scale Correction
- 8.10 Reduction Formulas
- 9 Software Licence Agreement
- 10 Glossary
- Appendix A Menu Tree
- Appendix B Directory Structure
- Appendix C Pin Assignments
•
Air temperature to 1 °C
•
Air pressure to 3 mbar
•
Relative humidity to 20 %
The air humidity influences the distance measurement if the climate is extremely hot and damp.
For high precision measurements, the relative humidity must be measured and entered along
with the air pressure and the temperature.
ppm
+5
+1
+2
+3
+4
+0
-10-20 0 10 20 30 40 50 °C
100%
80%
60%
40%
20%
TS_104
ppm Air humidity correction [mm/km]
% Relative humidity [%]
C° Air temperature [°C]
Type Index n Carrier wave [nm]
Combined EDM 1.0002863 658
The index n is calculated from the formula of the IAG Resolutions (1999), and is valid for:
Air pressure p: 1013.25 mbar
Air temperature t: 12 °C
Relative air humidity h: 60 %
Formula for visible red laser
· 10
x
4.126 · 10
-4
· h
-
0.29535 · p
ΔD
1
= 286.338 -
002419_002
DD
1
Atmospheric correction [ppm]
p Air pressure [mbar]
t Air temperature [°C]
h Relative humidity [%]
a
1
273.15
x (7.5 * t/(237.3 + t)) + 0.7857
If the basic value of 60 % relative humidity as used by the EDM is retained, the maximum possi-
ble error in the calculated atmospheric correction is 2 ppm, 2 mm/km.
Air humidity
Air humidity correction
Index n
Formulas
62 Technical Data