User Manual

Total Station Manual
31
item of PPM.
Ć
Ć
Ć
Ć
ƻ
5 After finishing setting,
press [F4](OK) to save settins
and back to previous
menu.Then press the key of
[F3](OK) to save the setting of
EDM and back to the function
of measurement.
[F4]
[F3]
Ć
1: The range of enter: Temp.(-30
ć
60
ć
), Press.(500hPa
1400hPa).
Ć
2: The instrument calculates the value of PPM according to the values of
temperature and pressure you enter.
Ć
3: Press [F1](PPM=0) can set the value of PPM to 0.
Ć
4: If instrument supports temperature pressure sensor, you can press [F2] to receive
the values of air pressure, temperature and calculate the correction valueautomatically.
2.3 Grid factor setting
When calculating the coordinates, the horizontal distance measured must
multiply by the scale factor.
Computation formula
1.Altitude factor=R/(R+ELEV)
R: The average radius of earth
ELEV: mean sea level altitude
2.Scale factor
Scale factor: Scale factor of the station
3.Grid factor
Grid factor=altitude factor×scale factor
Distance calculation
1. Grid distance
HDg=HD×grid factor
HDg: Grid distance