User guide

3
Signet 2759 pH/ORP Simulator and System Tester
SW2
Run
Cal
SW1
D1
+
S
-
BLK
RED
WHT
SHLD
Input from 2750
Output to
external equipment
4. Troubleshooting 2750 pH/ORP Electronics
Before using the 2759:
The most common cause of pH/ORP system problems is electrode depletion.
Perform routine electrode maintenance, including cleaning and inspection of the electrode, then calibrate the system.
See the electrode and 2750 Sensor manuals for detailed information.
If the problem persists, or to verify general system operation:
This test procedure requires the 3-2759.391 adapter cable.
Connecting the 2759 output to the 3-2759.391 adapter cable then connecting the adapter into the 2750 sensor electronics simulates the
output of the pH/ORP electrode.
Always use the HI button with the 3-2759.391 adapter cable.
Connect the 2759 to the 3-2759.391 adapter cable, then insert adapter into 2750 electronics.
Monitor the 2750 output using current monitoring device.
Slide the 2759 Mode selector switch to the proper position (pH or ORP).
Press output simulation buttons and then HI button.
(The HI button must be pressed after each output button.)
See Section 5: Response Chart for proper display.
Does the meter read a valid temp and pH/ORP?
Yes: The system is working ne or there is a problem with the electrode. Replace the electrode if necessary.
No: Problem is in 2750 Sensor Electronics. Replace the 2750.
Press any output simulation button to turn the 2759 on.
Press OFF button to turn the 2759 off.