User's Manual

Table Of Contents
Setup Requirements 163
Appendix A1
Installing RAPID Software
Setup Requirements
Set up your office to accommodate the new PillCam Capsule Endoscopy System. Review the following
workstation specifications:
Four electrical outlets are required to connect the following components: workstation computer,
monitor, printer, and one cradle. Each additional cradle requires an additional outlet.
RAPID Installation
Before installing RAPID on your computer, it is recommended to be aware of the following
prerequisites and other information:
Before installation, it is highly recommended to format the hard disk in NTFS format, if not already
done.
You need at least 1.5 GB free space on your computer to install RAPID.
For high resolution monitors, RAPID v8.3 offers the capability to resize and maximize the video
view.
֠
֠֠
֠
Note
You may use a Given approved power strip.
You may connect your monitor directly to the workstation using a DVI
cable (the DVI-VGA adaptor is not required).
!
Caution
Do not connect any component of the PillCam Capsule Endoscopy System to
the same outlet together with any appliance or device that has a high power
requirement (refrigerators, generators, devices with motors, etc.). When
setting up the system, make sure that the total power requirements for all of
the devices connected to a single outlet or circuit do not exceed the rated limit
for that circuit. If you are not sure of the rated limit, please consult your
maintenance department or an electrician.
Do not use a KVM (Keyboard, Video, Mouse) Switch with the PillCam
Capsule Endoscopy System.
֠
֠֠
֠
Note
Extra space is needed for air circulation and cable connectors behind the
workstation.