User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
Configuring RAPID Software
Regimen Manager 189
8. The regimens contain editable instructions in the Instructions for Patient column. Clicking the
instruction will prompt an editor. Information represented in brackets [ ] is generic and is intended
for completion by local staff with consideration to local names of the solution. Make any changes
and click OK to save the updated instruction.
9. Any other important information can be entered in the Additional Instructions box. This
information will appear in the patient printout.
10. When you are finished, click Save. This adds the regimen you created to the library of regimens
available.
11. By default, the post-capsule ingestion regimen includes at least the following alerts and
instructions:
• Alert 0: An instruction #0 (based on time passed from capsule ingestion) to take prokinetics to
facilitate passage of the PillCam COLON 2 capsule to the small bowel. Threshold of time
passed from capsule ingestion for alert 0 can be modified in the regimen manager for a specific
regimen.
•
Alert 1: An instruction #1 for taking laxative after detection by the PillCam recorder of PillCam
COLON 2 passage into the small bowel. It coincides with the activation of the AFR mode in the
PillCam COLON 2 capsule and the appearance of the AFR status icon in the right corner of the
status line at the top of the PillCam recorder DR3 display. The appearance of alert 1 can be
designated to appear either an additional 0 minutes or 15 minutes after original tentative alert 1
timing. Alert 0 will not occur if alert 1 was raised before it.
• The End of Procedure (EOP) alert occurs when no more capsule images are received by the
PillCam recorder or enough time has elapsed in the procedure. This may occur several minutes
after capsule excretion or after the capsule battery was depleted without excretion in a long
procedure.
In general, additional alerts and instructions can be inserted between alert 1 and the EOP alert to
reflect the physician's preference for post-ingestion instructions in a PillCam COLON 2 procedure.
These instructions can be entered and edited, saved and approved through the Regimen Manager
and be made ready to be selected in the Check-in for a PillCam COLON 2 procedure.
֠
֠֠
֠
Note
Alert times are designated by the PillCam recorder and the RAPID
software. In rare cases, short delays may occur.