User's Manual
Table Of Contents
- User Manual
- RAPID® v8.3
- Table of Contents
- Chapter 1
- Using This Guide
- Chapter 2
- Indications, Contraindications, Warnings, Cautions
- Chapter 3
- Welcome to PillCam Capsule Endoscopy
- Chapter 4
- Preparing for PillCam Capsule Endoscopy
- Preparing the Patient
- Preparing the Required Equipment
- Creating Patient Instructions for the Procedure
- General Patient Guidelines During the Procedure
- Performing Patient Check-in
- Fitting Equipment on the Patient
- Chapter 5
- Know Your PillCam Recorder
- Chapter 6
- Performing PillCam Capsule Endoscopy
- Chapter 7
- Creating RAPID Videos
- Chapter 8
- Reviewing and Interpreting RAPID Videos
- Loading a Study with the Study Manager
- Using the Study Manager
- Overview of the RAPID Interface
- Using RAPID to View a Video
- RAPID Video Files
- Working with Findings
- Creating a PillCam Capsule Endoscopy Report
- RAPID Atlas
- Lewis Score
- Generating a Report
- Appendix A1
- Installing RAPID Software
- Appendix A2
- Configuring RAPID Software
- Appendix A3
- PillCam Equipment Maintenance
- Appendix A4
- Troubleshooting
- Appendix A5
- Technical Description
- System Labeling
- Essential Performance
- Warnings
- Cautions
- System Specifications
- PillCam SB 2 Capsule
- PillCam SB 3 Capsule
- PillCam UGI Capsule
- PillCam COLON 2 Capsule
- Sensor Array PillCam Recorder DR2
- Sensor Array PillCam Recorder DR3
- PillCam Recorder DR2/DR2C
- Cradle PillCam Recorder DR2
- PillCam Recorder DR3
- PillCam Recorder DR3 SDHC Memory Card
- Cradle PillCam Recorder DR3
- DC Power Supply
- RAPID for PillCam Software
- Guidance and Manufacturer's Declarations
- Index
PillCam Equipment Maintenance
PillCam Sensor Cleaning 205
PillCam Sensor Cleaning
This section includes instructions for cleaning the PillCam equipment.
Cleaning the PillCam Sensor Belt
Cleaning the PillCam Sensor Array
For mild cleaning (dirt, sweat), wipe the sensors gently with alcohol wipes (up to 70%). The alcohol
will not remove the adhesive. Use alcohol sparingly and allow the sensor array to dry for 20 minutes.
To remove adhesive from the sensor array (not from the human body), use white benzene.
Alternatively, use one of the following medical adhesive removers to remove adhesive:
• B-508 Secure Solvent
• B-202 Hollister Solvent
• B-206 Detachol Adhesive Remover
Use all precautions as defined by the manufacturer.
Cleaning the Recorder Pouch
To clean the recorder pouch, wipe down all surfaces with alcohol (70% isopropyl or ethyl alcohol)
making sure that all surfaces are exposed to alcohol for at least 1 minute.
֠
֠֠
֠
Note
To clean the PillCam equipment surfaces, we recommend using alcohol
wipes (up to 70%). Usage of other classes of disinfectants (Aldehydes,
Oxidizing Agents, Quaternary Ammonium, Chlorine compounds, Iodophor,
Phenolic compounds) may stain or damage the materials.
֠
֠֠
֠
Note
Refer to the product insert supplied with your sensor belt for full instructions
on fitting, usage, cleaning, and technical description.
!
Warning
Use white benzene only in a well ventilated area according to all
precautions and instructions on the label.