74 PillCam Recorder DR2
PillCam Recorder DR2
General
The PillCam recorder DR2 battery is limited to 350 use cycles. Contact customer service for a
replacement battery when necessary.
Turning On and Off
This illustration shows the front and back view of the PillCam recorder DR2:
The On/Off button is on the top left side on the back of the PillCam recorder DR2:
To turn on, press and hold the on/off button until you hear a long beep followed by a short one and the
LEDs start flashing.
To turn off, press and hold the On/Off button until you hear a beep and the LEDs turn off.
֠
֠֠
֠
Note
Once the number of uses exceeds 350, the following message appears (in
the log file only): DR2 battery must be replaced before further use. Contact
customer support to arrange replacement.
֠
֠֠
֠
Note
Automatic shutdown: After the PillCam recorder DR2 has been checked
in, it goes into standby mode ready to receive capsule signals when
removed from its cradle and starts recording as soon as a signal is received
from a transmitting capsule. If after removal from the cradle no signal is
received for 90 minutes, the PillCam recorder DR2 automatically shuts
down.
After starting to receive signals from a capsule, if there is a gap of 60
minutes of no signal reception, the recorder shuts down.
Locking
handle in
open position
PillCam
recorder DR2
LED
Capsule LED
ON/OFF button
Battery
LEDs