70 PillCam Recorder DR3
Battery and Capsule Icons
The battery icon on the left side of the status line at the top of the display indicates the status of the battery
in 10% increments. These icons appear in the top status line of the PillCam recorder DR3 screen.
PillCam Recorder DR3 LEDs
The status indications of the LED indicators on the top of the PillCam recorder DR3 for the most common
PillCam recorder events are as follows:
Icon Battery Status Icon Capsule Reception Status
Battery fully charged Signal weak, recording with
noise
Battery charge level at 10%
intervals
Signal strong, recording with
noise
Battery empty, PillCam
recorder shuts down
Signal weak, but recording OK
Battery charging Signal strong, and recording
OK
LEDs PillCam Recorder DR3 Status
CHECK-IN
Blinking in white
PillCam recorder DR3 is being checked in.
Blinking rate = very fast.
Constant on in white
PillCam recorder DR3 is checked in and ready to receive
capsule signals.
Blinking in white
PillCam recorder DR3 is checked in and receiving signals
from an unpaired capsule.
Blinking rate = capsule frame rate.
RECORD
Blinking in blue at capsule
frame rate
PillCam recorder DR3 is receiving paired capsule signals and
recording.
Blinking rate = capsule frame rate.
LEDs off
Recording done and raw data is available for downloading.
DOWNLOAD
Blinking in green
PillCam recorder DR3 has started downloading.
Constant on in green
PillCam recorder DR3 has completed downloading.