Appendix: Electrical Safety
4. Click OK. You will be asked to confirm the clinic identification number you
entered; click Yes to proceed to the First Login Sequence screen.
5. Click OK. The administrator user login screen appears.
6. Create a username and password for the administrator account for MotiliGI. The
username and password must contain a minimum of 5 alphanumeric characters.
7. Verify the password by re-entering it, and click Login. You will be required to
enter this username and password each time you start the program. The MotiliGI
Start-Up screen appears.
Connecting the Docking Station to the System Computer
1. Connect the docking station to an AC power source using the power cord.
2. Connect the docking station to the system computer using the USB cable.
3. Turn the data receiver on by depressing the event button.
4. Dock the data receiver by placing it in the docking station cradle. Push down on
the top of the data receiver to ensure the data receiver is fully docked.
5. Confirm the system computer and docking station are communicating by noting
the presence of the data receiver connected icon in the MotiliGI status bar (the
horizontal bar at the bottom of the MotiliGI user interface).
x If you observe the receiver communication disabled and test state unknown
icons, either the docking station is not properly connected to the system
computer, or the data receiver is not fully docked.
Reseat the data receiver in the docking station and ensure the docking station is
connected to both the computer and a power supply. The test state unknown icon will
disappear once the connection is established. If it does not disappear, contact Technical
Support for assistance.
6. Before starting a test, fully charge the Data Receiver. See Chapter 4: Charging the
Data Receiver for instructions.