User Manual

Appendix: Electrical Safety
6WDUWLQJD7HVW
Preparing the System
Perform these steps immediately before or after a patient’s arrival at the office.
1. Turn on the system computer.
2. Open MotiliGI:
x Click on the desktop.
x Select Start > All Programs > SmartPill > MotiliGI.
3. If prompted, enter a username and password. Click OK. The main startup screen
appears.
Performing the Test ± Test Initiation Wizard
Step 1: Connect the Data Receiver
Caution
Do not close MotiliGI while operating the Test Initiation Wizard because data can be lost. Complete or
cancel the Wizard first.
Caution
You may cancel the wizard any time. However, a capsule exposed to the pH calibration buffer must be
ingested within 2 hours or discarded.
1. Follow the prompts. Check the results:
Component
Success
Troubleshooting
Docking station’s
LED
Green
x Red
The data receiver may not be fully charged. See Chapter 4
:
Charging the Data Receiver.
x Yellow
Undock and re-dock the data receiver in the docking
station. If the yellow light persists, contact SmartPill
Technical Support.
Data receiver
display
The “Looking For
Capsule” m
essage
appears.
x LCD is blank
Turn on the data receiver by pressing the event button.
System
Computer
x 2 beeps
x These icons
appear in the
lower-right
corner of the
screen:
x
x No beeps
x Check the computer’s sound level.
x Undock and re-dock the data receiver in
the docking
station. If there are still no beeps, contact Technical
Support.
x The “Receiver is Not Fully Charged” message appears.
The charge is insufficient for a 5-day test.