User Manual

Appendix: Electrical Safety
Step 4: Activate Capsule
1. Follow the prompts.
2.
Figure 4 - Capsule Activation Window
Caution
Once the capsule is activated, keep the capsule at least one (1) foot away from the activation fixture. If the
capsule is kept near the activation fixture, stray magnetic fields from the activation fixture may deactivate the
capsule.
3. Confirm the capsule has activated:
x Observe the data receiver’s display for at least 1 minute. The message “Locked
onto Capsule” and the capsule’s serial number will appear on the data receiver
for approximately 4 seconds.