User Manual

Appendix: Electrical Safety
x Calibration normally takes less than 3 minutes, but may take up to 15 minutes.
x During calibration, it is normal for fluctuations in the pH value to occur.
Calibration is complete when all these signals appear:
x The pH value stabilizes at 6.0.
x The “pH Calibration Complete” message appears.
x The MotiliGI status bar icon indicates the capsule is ready for use and the
data receiver is recording information from the capsule.
If the capsule does not calibrate within 15 minutes, this message appears: “pH
Calibration Timeout.”
1. Ensure the capsule is immersed in buffer.
2. Click Back.
3. Click Next. The calibration process reinitiates.
x If the calibration fails after 2 attempts contact Technical Support.
Step 9: Ingestion
1. Follow the prompts. Record the time of completion in the patient diary.
Caution
The meal must be eaten immediately before capsule ingestion.