7. USER PARAMETERS
1. When the scale is in normal weighing mode, press and hold the key and key for
3 seconds until ‘SEtUP is shown on the display.
2. In the SETUP mode, press the key to change the flashing digits, press the key to shift the
flashing data entry position from left to right, and press key to confirm and move to the
next parameter setting. Press the key to exit the set up mode.
Parameters setting summary:
Parameter
x/xy
Remark
244241
Setting
244242
Setting
P1.xy
00-15
Auto-off time:
No auto-off = 00.
01-15 minutes auto-off time.
05
05
P2.xy
00,01,02
0 = Only Hold Function
1 = Only Print Function
2 = both HOLD and PRINT function (pressed down less
3s,this key works as Print function; pressed down
more than 3s,this key works as HOLD function)
02
02
P3.xy
00-50
Hold function:
0 = disable hold function
1 = hold larger weight reading
2-50 = hold reading when the variety is within +2~50d,
auto release hold function when weight is below 10d and
auto-hold new stable weight (more than 10d)
02
02
P4.x
0-3
0 = No RS232 and USB Function.
1 = Press PRINT key to output display data when scale is
stable
2 = Press PRINT key to output gross, TARE and net
weight when scale become stable;
3 = Continuously outputs display data.
4 = continuously output gross, TARE and net weight;
5 = Output display data one time when scale is stable;
6 = Output gross, TARE and net weight one time when
scale become stable
7 = Bi-directional communication (the scale receives and
executes commands from the HOST device)
2
2
P5.x
0-4
Communication baud rate:
0=1200bps,
1=2400bps,
2=4800bps,
3=9600bps,
4=19200bps
3
3