6.3 Phantom
<SAM Twin Phantom>
Shell Thickness
2 ± 0.2 mm;
Center ear point: 6 ± 0.2 mm
Filling Volume
Approx. 25 liters
Dimensions
Length: 1000 mm; Width: 500 mm; Height:
adjustable feet
Measurement Areas
Left Hand, Right Hand, Flat Phantom
The bottom plate contains three pair of bolts for locking the device holder. The device holder positions are adjusted to the
standard measurement positions in the three sections. A white cover is provided to tap the phantom during off-periods to
prevent water evaporation and changes in the liquid parameters. On the phantom top, three reference markers are
provided to identify the phantom position with respect to the robot.
<ELI Phantom>
Shell Thickness
2 ± 0.2 mm (sagging: <1%)
Filling Volume
Approx. 30 liters
Dimensions
Major ellipse axis: 600 mm
Minor axis: 400 mm
The ELI phantom is intended for compliance testing of handheld and body-mounted wireless devices in the frequency