7.2 Power Reference Measurement
The Power Reference Measurement and Power Drift Measurements are for monitoring the power drift of the device under
test in the batch process. The minimum distance of probe sensors to surface determines the closest measurement point
to phantom surface. This distance cannot be smaller than the distance of sensor calibration points to probe tip as defined
in the probe properties.
7.3 Area Scan
The area scan is used as a fast scan in two dimensions to find the area of high field values, before doing a fine
measurement around the hot spot. The sophisticated interpolation routines implemented in DASY software can find the
maximum found in the scanned area, within a range of the global maximum. The range (in dB0 is specified in the
standards for compliance testing. For example, a 2 dB range is required in IEEE standard 1528 and IEC 62209 standards,
whereby 3 dB is a requirement when compliance is assessed in accordance with the ARIB standard (Japan), if only one
zoom scan follows the area scan, then only the absolute maximum will be taken as reference. For cases where multiple
maximums are detected, the number of zoom scans has to be increased accordingly.
Area scan parameters extracted from FCC KDB 865664 D01v01r04 SAR measurement 100 MHz to 6 GHz.