Operating Instructions SECUTEST®SII Test Instrument for DIN VDE 0701, 0702 and 0751 3-349-223-03 8/11.
10 9 8 7 6 RS232 5 4 SONDE 5 4 Feature F01 3 L 3 2 N 2 L VDE 0701 11 VDE T. 240 0702 T. 200 T. 1 N VDE 0751 12 OFF 13 1 SL 1 MENU FUNCTION-TEST 230 V SECUTEST SII 14 15 ☞ 23 20 18 19 18 17 16 22 Note Contact problems with exposed conductive parts when using the standard probe with test tip In order to assure good contact, surface coatings must be removed from devices under test with special tools at a suitable location so that the surface has a metallic shine.
Upper left figure 1 Jack for protective conductor at device under test (feature F01) 2 Jack for neutral conductor at device under test (feature F01) 3 Jack for phase conductor at device under test (feature F01) 4 Jack for connecting the probe 5 Jack for connecting the probe 6 Function selector switch – VDE-…: Automatic test sequence according to selected standard – OFF: Instrument is switched off (no disconnection from mains) – MENU: Setup: Device configuration RSL… : Individual tests – FUNCTION TEST: Funct
Contents Page Contents Page 1 Applications ...................................................................................6 9 Individual Measurements ............................................................16 1.1 1.2 1.3 1.4 Table: Types of DUTs – Tests – Regulations ....................................................6 Table: Individual Measurements and Regulations ............................................6 Table: Leakage Current Types ...................................................
Contents Page Contents Page 12 Storage in PSI Module (Feature E01) and Database Operations (Feature KB01 or DBmed option) ........ 38 19 Maintenance - Calibration ........................................................... 46 12.1 12.2 12.2.1 12.2.2 Storing Measurement Data in the PSI Module ................................................38 Database Operations ...................................................................................38 Setting Up Test Sequences with PC Software ..........
• • • 1.2 Attention! The test instrument may not be used for measurements within electrical systems! Table: Individual Measurements and Regulations 0.
1.
2 Safety Features and Precautions The SECUTEST®SII test instrument has been manufactured and tested in accordance with the following safety regulations: IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1 and DIN VDE 0404, DIN VDE 0413 Part 2 and 4. When used for its intended purpose, the safety of the user, the test instrument and the device under test (electrical equipment or electrical medical device) is assured.
3 Initial Start-Up 3.1 Connection to the Mains (230 V, 50 Hz) U L – N = 230 V ➭ Connect the mains plug at the test instrument to the mains power outlet. The function selector switch can be set to any position. If a mains outlet (earthing contact outlet) is not available, or if only a 3-phase outlet is available, the adapter socket can be used to connect the phase conductor, the neutral conductor and the protective conductor.
3.2 Automatic Recognition of Mains Connection Errors The device automatically recognizes mains connection errors if the conditions in the following table have been fulfilled. The user is informed of the type of error, and all measuring functions are disabled in the event of danger.
4.3 Adjusting Contrast MENU Set the selector switch to MENU. Select the Setup menu, “back” is highlighted. Activate contrast adjustment. Press and hold the key. Adjust contrast. Return to the menu. Store the contrast setting to permanent memory with the save function in the setup menu. GOSSEN METRAWATT GMBH 4.
5 Classification of Devices Under Test 5.1 Safety Classes Devices assigned to all of the following safety classes are equipped with basic insulation, and provide for protection against electrical shock by means of various additional precautions as well. Safety Class I Devices Exposed, conductive parts are connected to the protective conductor so that they are not charged with voltage if the basic insulation should fail.
6 Abbreviations AE B, BF, CF OE ΔI ΔImax EDLA1/A2 Error condition: application part grounded Classifications for application parts Operational earth Differential current, residual current Maximum residual current (during function test) Equivalent device leakage current with designation A1/A2 (cross-reference in the standard) EDLPX±PE Equivalent device leakage current for portable x-ray +PE: with additional PE conductor –PE: without additional PE conductor EDLSCII Equivalent device leakage current for dev
7 Connecting the Device Under Test ➭ Connect the DUT in accordance with the schematic diagrams included in the online help function.
8 Configuring Device Parameters MENU General device parameters can be configured and saved using Setup with the selector switch in the MENU position. Select Setup/Menu and acknowledge. Select parameter and acknowledge, change setting and acknowledge. Settings x / – … = function activated / deactivated Limit values ... Illumination Background illumination for the LCD.
9 Individual Measurements MENU 9.1 Measuring Protective Conductor Resistance Resistance is measured: • Between each exposed conductive part of the housing and the earthing contacts at the mains and the device plug (if a removable mains connector cable is used), or the protective conductor terminal for permanently installed devices.
Combined Test – Differential Protective Conductor Resistance Zero balancing is also possible for protective conductor measurement. With zero balancing, all subsequent measurements are adjusted with an offset such that 0 Ω is displayed for a selected reference point which is connected to the protective conductor.
Minimum Allowable Limit Values for Insulation Resistance Sequence ☞ Note All switches at the device under test must be set to the on position during measurement of insulation resistance, including temperature controlled switches and temperature regulators as well. Measurement must be performed in all program stages for devices equipped with program controllers. Test Standard Test Voltage RISO SC I SC II SC III Heat VDE 0701 Part 1:2000 1 MΩ 2 MΩ 0.25 MΩ 0.3 MΩ ∗ VDE 0702: 2004 1 MΩ 2 MΩ 0.
9.3 ! Measuring Leakage Current Attention! During leakage current measurement it is essential to ensure that the DUT is operated at mains voltage. Exposed conductive parts may be subject to hazardous contact voltage during the test and may not be touched under any cricumstances. (A mains shutdown takes place when the leakage current exceeds approx. 10 mA). Select the Ixx measurement and start. 9.3.
9.4 Maximum Allowable Limit Values for Equivalent Leakage Current in mA Equivalent Leakage Current General Measurement of equivalent leakage current is required for: • DIN VDE 0701 and 0702, after passing the insulation L test mA Test Standard IEL VDE 0701-1: 2000 SC I: 3.5 1 mA/kW 1) SC II: 0.5 VDE 0702: 2004 3.5 1 mA/kW 1) SC II 0.
Sequence Current, which would flow during leakage current measurement conducted in accordance with device regulations and with nominal line voltage, is displayed during this type of equivalent leakage current measurement. Leakage current measurement in accordance with the respective device regulations is usually not possible, because the device would have to be set up in an electrically isolated fashion, or connected to an earth isolated power supply to this end. 9.5 Probe Voltage Uprobe – Max.
9.6 Alternating / Direct Voltage UAC/DC – Max. 253 V (feature F01) Direct, alternating and pulsating voltages of up to 253 V can be measured between the N (2) and L (3) connector jacks. 9.8 Measurements with Accessories (feature F01) 9.8.1 Alternating Current IZ with Clip-On Transformer (feature F01) Connection Select the UAC/DC measurement and start. 9.7 Resistance R (feature F01) Resistance of up to 150 kΩ can be measured between the PE (1) and N (2) connector jacks.
9.8.2 Temperature T with Pt100/1000 Sensor (feature F01) Connection Zero Balancing Sensor cable resistance can be compensated with this function: ➭ Short circuit the ends of the sensor leads and determine resistance as shown below. Zero Balancing The determined value can be stored directly ( key) or changed first. The data entry menu is opened with the key.
10 Function Test The device under test can be subjected to a function test with line voltage via the integrated test socket. Function Test In addition to testing with the selector switch in the function test position, or with the Function parameter in the MENU position (feature F01 only), a function test can also be performed immediately after safety testing has been passed in accordance with the selected standard (not possible for safety class III devices).
Short-Circuit Test 1 Test for shorts between conductors N and L. 2 Test to determine whether or not the N or L conductors are shortcircuited to the protective conductor. The test socket can be rendered voltage-free with the key (14), or the function test can be ended with the key (12). ☞ Note A short-circuit at the device under test is recognized automatically by the test instrument. A message appears at the display (9), and the function test is disabled.
11 Measurements – Selector Switch Settings for Standards If measurements need to be performed in accordance with given standards which require specific tests, and if results need to be documented with a test report, an automatic test sequence is advisable instead of individual measurements. ➭ Connect the SECUTEST®SII to mains power. A mains connection test is initialized (see chapter 3.2 on page 10). ➭ Connect the DUT to the test socket at the SECUTEST®SII (see chapter 7 on page 14).
11.2 Setting Up Test Sequences All possible sequence settings for all of the regulations are listed below. Select the Sequence ... menu and acknowledge. Select the Setup... menu from the initial program window and acknowledge Repair tests, periodic testing Save All of the settings in the setup menu, i.e. configuration of measuring parameters and current limit values, can be saved with this command.
11.
Test Sequence per VDE 0701 Select connection, select test regulation VDE 0701 part 1 or 200, classify the DUT (SC I, II or III), part 1: extension cable X/– (with/without) Visual inspection OK? Yes/No Yes Test OK? Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under Sequence ... Switch DUT on Short-circuit at DUT? No No Safety Class I only: Contact all exposed, conductive parts with the probe.
11.5 Testing Devices in Accordance with DIN VDE 0701, Part 240 Testing safety class I and II data processing equipment and office machines as individual devices and in combination with one another.
Test Sequence per VDE 0701, Part 240 Select connection, select test regulation VDE 0701, part 240, classify the DUT (SC I, II or III), combination testing X/– (yes/no) Visual inspection OK? Yes/No Yes Test OK? Switch DUT on Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under Sequence ... Short-circuit at DUT? No No SC I: RPE OK? Contact all exposed, conductive parts with the probe.
11.6 Testing Devices in Accordance with DIN VDE 0702:2004 The following measurements can be performed in accordance with the above mentioned standard: • Protective conductor RPE (permanent connection or with plug) Test current: DC ±200 mA • Insulation measurement RISO (can be deactivated, e.g.
Test Sequence per VDE 0702:2004 Select connection, select test regulation VDE 0702, classify DUT (SC I, II or III), extension cable X/– (with/without) Visual inspection OK? Yes/No Yes Test OK? Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under Sequence ...
11.7 Testing Extension Cables for VDE 0701, Part 1, and VDE 0702 (option: EL1 adapter) Extension Cables up to 5 m Long Protective conductor resistance between the earthing contact at the mains plug and all exposed metal parts may not exceed 0.3 Ω for safety class I devices. This value may not exceed 1 Ω for permanently connected data processing systems (DIN VDE 0701, part 240). Extension and Connector Cables Longer than 5 m Per DIN VDE 0702, an additional cable resistance of 0.
11.8 Testing Multiple Outlets for VDE 0702 (optional EL1 adapter) ➭ up to version 5.10: Do not select “WITH EL1” in the initial menu. The following must appear in the line: Ext. Cable – WITH EL1. “Manual sequence” must be selected. ➭ from version 5.11 onwards: Select “X WITH EL1” in the initial menu. The following must appear in the line: Ext. Cable X WITH EL1. “Manual sequence” must be selected. ➭ A visual inspection must always be performed.
11.
Test Sequence per VDE 0751 Select connection, select test regulation VDE 0751, classify the DUT (SC I or II), application part? (type B/BF/CF) Test OK? Visual inspection OK? Yes/No Yes Switch DUT on Framed with dashed line: The test is only run – If it has been activated in the initial program window, or in the Setup menu under Sequence ...
12 Storage in PSI Module (Feature E01) and Database Operations (Feature KB01 or DBmed option) 12.1 Storing Measurement Data in the PSI Module Upon completion of a measurement – „Test passed/Test failed“ is shown on the display – you can save the measurement data to the memory of the PSI module. ➭ Press the STORE key at the PSI module to this end. A text entry field is shown. ➭ You may now enter your comments on the measurement and/or an identification number.
14 Printing via PSI Module (Feature E01) or with the Printer Adapter (DA-II accessory) 14.1 Print Out Test Data at the PSI Module Upon completion of a measurement – „Test passed/Test failed“ is shown on the display – you can print out the measurement data via your PSI module. ➭ Press the PRINT key at the PSI module to this end. The print menue is displayed. ➭ Select „Protocol“ with your cursor and acknowledge with . ➭ Press the PRINT key once more to start the print-out.
16 Characteristic Values Measuring Measuring Range / Resolu- Nominal Open-Cir- Nominal Short- Internal Reference cuit Volt- Current Circuit ResisResist. Error Nominal Range of tion Voltage RREF age U0 IN Cur. IK tance RI Use UN 0.000 … 2.100 Ω 1 mΩ 4.5 … 9 V >200 mA ±(5% rdg. + 10 d) Device Protective — — — — DC DC > 10 d Conductor 2.11 … 31.00 Ω 10 mΩ Resistance RPE 0.050 … 1.500 MΩ 1 kΩ ±(5% rdg. +10 d) 50 … 500 1.0 • UN … > 1mA < 10 mA — — Insulation Resistance 1.01 … 10.00 MΩ 10 kΩ 1.
Multimeter Measurements for Devices with Feature F01 Function UAC/DC Measured Quantity Measuring Range / Nominal Range of Use Resolution OpenCircuit Voltage U0 0 ... 253.0 V , and 0.1 V — — 0 ... 150.0 kΩ Voltage Low Voltage, SC III R Resistance IClip Current with Clip-On CurrentVoltage Transformer WZ12C Temp.
Power Supply Line Voltage Line Frequency Power Consumption for function test Electrical Safety Safety Class Nominal Voltage Test Voltage Measuring Category Fouling Factor Safety Shutdown 207 V … 253 V 50/60 Hz Approx. 30 VA Continuous max. 3600 VA, power is conducted through the instrument only, switching capacity ≤ 16 A I per IEC 61010-1/EN 61010-1/VDE 0411-1 230 V 3.
17 RS 232 Interface The RS 232 port is used to connect the SECUTEST®PSI module (accessory), which can be installed to the lid of the SECUTEST®SII, to a PC or a barcode scanner. 17.
18 Appendix 18.1 Evaluation of Measured Values for Individual Measurements as well as for Calculated Quantities In order to assure that the limit values for the individual measurements are always observed, device measuring error must be taken into consideration.
18.2 Index I S A Illumination ..............................................................15 Incl. Service Error ....................................................15 Individual Measurements ..........................................16 Insulation Resistance ...........................................6, 17 Insulation Resistance Limit Values ............................18 Interface ..................................................................43 IT Network ..........................................
19 Maintenance - Calibration Housing Maintenance No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives or solvents. Calibration According to the new DIN VDE 0702:2004-06 standard the following applies as from 1/6/2004: „Measuring instruments used for periodic testing must be inspected and calibrated at regular intervals.
Competent Partner GOSSEN METRAWATT GMBH is certified in accordance with DIN EN ISO 9001:2000. Our DKD calibration lab is accredited by the Physikalisch-Technischen Bundesanstalt (German Federal Institute of Physics and Metrology) and the Deutscher Kalibrierdienst (German Calibration Service) in accordance with DIN EN ISO/IEC 17025 under registration number DKD–K–19701.
Edited in Germany • Subject to change without notice • PDF version available from the Internet GOSSEN METRAWATT GMBH Thomas-Mann-Str. 16-20 90471 Nürnberg • Germany Phone: +49-(0)-911-8602-0 Fax: +49-(0)-911-8602-669 E-mail: info@gossenmetrawatt.com www.gossenmetrawatt.