11.6 Testing Devices in Accordance with DIN VDE 0702:2004
The following measurements can be performed in accordance with the
above mentioned standard:
Protective conductor R
PE
(permanent connection or with plug)
Test current: DC ±200 mA
Insulation measurement R
ISO
(can be deactivated,
e.g. if danger exists that voltage sensitive components in data
processing systems might be damaged) plus equivalent leakage
current
or
Contact current for safety class II
or
Differential current
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
14 for other types of connection.
Class If the test instrument is equipped with features B00,
01 or 03, and if the DUT has been connected to the
test socket, safety class recognition is performed
(SC I or SC II). Otherwise, the safety class must be
specified manually.
Ext. Cable x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the
EL1 adapter (optional), either separately or in
combination with a device.
ISO-R x: Insulation resistance measurement is performed.
ID No. See parameters database in chapter 11.2 on page
27.
Setup See chapter 11.2 on page 27 regarding setup of the
measuring sequence.