3.12 CV Mode Calibration
In this mode, three voltage points: 0.5 V, 22 V, 34.5 V (take LABKON P500 35 V / 14.5 A, K148A as example) should be
calibrated. A complete overview of calibration checkpoints for all LABKON models is listed on page 57 of this manual.
3.12.1 Wiring
3.12.2 0.5 V Calibration
Press “Enter” key, you will see the following figure in the display:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press the “Clear” key to
remove the wrong input value of current digit. Then press the “Enter” key to confirm.
3.12.3 22 V Calibration
Press “Enter” key to verify and then you will see the followings:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press the “Clear” key to
remove the wrong input value of current digit. Then press the “Enter” key to confirm.