User manual

SSL ELECTRONIC LOAD SERIES 32 EL
14
GMC-I Messtechnik GmbH
2.3.1 Setting the maximum current
This parameter defines the upper limit for the current setting range. When you select the Maximum Current function, the
LCD will display:
Max Curr=****A
New=
Set the maximum current value by using the number
keyboard or the rotary knob, then press “OK” for
confirmation.
2.3.2 Setting the maximum power
This parameter defines the upper limit for the power setting range. When you select the Maximum Power function, the
LCD will display:
Max Power=****W
New=
Set the maximum power value by using the number
keyboard or the rotary knob, then press “OK” for
confirmation.
2.3.3 Setting Minimum Input Voltage
This function is used to set the minimum input voltage, when the input voltage is lower than this value, the SSL
KONSTANTER 32 EL electronic load will automatically stop. This function is used to discharge batteries. When the
battery’s voltage drops down to the set value, the load switches off automatically, keeping the battery from over-discharge.
When you select Min Input Volt function, the LCD will display:
Min Volt=****V
New=
Set the minimum input voltage by using the number
keyboard or the rotary knob, then press “OK” for
confirmation.
2.3.4 Setting Baud Rate
This function is used to set the communication baud rate.
When you select the Baud Rate function, the LCD will display:
4800 bps
9600 bps
19200 bps
38400 bps Def*
Select the Baud Rate by using the UP and DOWN keys
or the rotary knob, then press “OK” for confirmation.