Manual
Table Of Contents
- 1 Safety Instructions
- 2 Applications
- 3 Documentation
- 4 Operating Overview
- 5 Initial Startup
- 6 System Settings
- 6.1 Digital Keyboard
- 6.2 Firmware
- 6.3 Language Selection
- 6.4 Time and Date Settings
- 6.5 Digital Display Settings
- 6.6 Set Display of Leading Zeros
- 6.7 Configuring Password Protection
- 6.8 Viewing the Instrument’s Name
- 6.9 Installing a Functions Expansion
- 6.10 Switching the Instrument Off
- 6.11 Default Settings (instrument reset)
- 7 Control Functions
- 8 Measurements
- 8.1 Enabling Parameter Changes
- 8.2 Insulation Resistance Measurement – RISO Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.3 Inter-Turn Short Circuit Measurement – COIL Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.4 Absorption Index Measurement – DAR (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.5 Polarization Index Measurement – PI (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6 Voltage Measurement
- 8.6.1 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter
- 8.6.2 Duty Cycle Measurement – Duty AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.3 RPM Measurement – RPM AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.4 Direct and Pulsating Voltage Measurement, V DC and V (AC+DC)
- 8.7 Resistance Measurement “W”
- 8.8 Capacitance Measurement F
- 8.9 Temperature Measurement with Resistance Thermometers – Temp RTD
- 8.10 Temperature Measurement with Thermocouple – Temp TC
- 8.11 Continuity Test
- 8.12 Diode Testing with Constant Current of 1 mA
- 8.13 Milliohm Measurement – Rlo (2-wire measurement) (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.14 Milliohm Measurement – mW/4 (4-wire measurement)
- 8.15 Current Measurement
- 8.15.1 Direct and Pulsating Current Measurement, Direct Connection, – A DC and A (AC+DC)
- 8.15.2 Alternating Current and Frequency Measurement, Direct Connection, – AAC and Hz
- 8.15.3 Direct and Pulsating Current Measurement with Current Clamp Sensor – ADC und A (AC+DC)
- 8.15.4 Alternating Current Measurement with Current Clamp Sensor – AAC and Hz
- 8.16 Measuring Sequences – Test Sequences
- 9 Interface Operation and Software
- 10 Characteristic Values
- 11 Maintenance and Calibration
- 12 Accessories
- 13 Returns and Environmentally Sound Disposal
- 14 CE Declaration
- 15 Addresses
Gossen Metrawatt GmbH 27
8.3.1.1 Preparing for Measurement
Note
Inter-turn short circuit measurements may only be con-
ducted on voltage-free coils.
➭ Set the rotary switch to “
RISO
” or “
COIL
”.
➭ Repeatedly press the Func softkey until the measurement view
for COIL appears at the display.
➭ Select the motor or coil type to be tested under the measure-
ment type parameter in “Setup for present measurement” (see
below).
Selecting the Type of Measurement
> Momentary Measurement Setup > Parameter
Measurement Type
Coil / 1-Ph. Motor / 3-Ph. Motor
3 x > Measurement View
1-Ph. Motor: AC motor (L1)
3-Ph. Motor: 3-Phase motor (U-V, U-W, V-W)
Coil Motor with up to 15 coils (L1 - L15)
➭ Select polarity: unipolar or bipolar (see below).
Polarity Selection
> Momentary Measurement Setup > Parameter
Polarity
Bipolar / Unipolar
3 x > Measurement View
S– A
V
Tem p
m/4
COIL
S+
0V!
!
V
U
W
Example
COIL
COIL
TEST ADAPTER
with Star Connection
M
Connect
COIL TEST ADAPTER
and measurement cables
U
Surge max
1200 V!
RISO
COIL
V
U
W
L1
L3
L2
Example
with Delta Connection
Pulse
Current
up to 600 A
is Possible!
MENU
OK
OK
ESC
MENU
OK
OK
ESC
Func
Func
Func
Zero:
RISO
COIL
RISO
Coil
DAR
PI
The above symbol
blinks when test volt-
age is active.
Func