Instructions
56 GMC-I Messtechnik GmbH
14.4 Leakage Current Measurement 
with PRO-AB Leakage Current Adapter as Accessory 
– I
L 
Function (PROFITEST MXTRA & SECULIFE IP only)
Applications
Measurement of contact voltage in accordance with DIN VDE 
0107, part 10, as well as continuous leakage and patient auxiliary 
current per IEC 62353 (VDE 0750, part 1) / IEC 601-1 / EN 60 
601-1:2006 (Medical electrical equipment – General requirements 
for basic safety), is possible using the PRO-AB PRO-AB leakage 
current measuring adapter as an accessory with the 
PROFITEST MXTRA test instrument.
As specified in the standards listed above, current values of up to 
10 mA may be measured with this measuring adapter. In order to 
be able to fully cover this measuring range using the measure-
ment input provided on the test instrument (2-pole current clamp 
input), the measuring instrument is equipped with range switching 
between transformation ratios of 10:1 and 1:1. In the 10:1 range, 
voltage dividing takes place at the same ratio.
Connection and Test Set-Up
In order to perform the leakage current measurement, the 
adapter’s measurement outputs must be plugged into the mea-
surement inputs at the left-hand side of the PROFITEST MXTRA (2-
pole current clamp input and probe input).
Either of the leakage current measuring adapter’s inputs is con-
nected to reference earth (e.g. safe earth electrode / equipotential 
bonding) via a measurement cable. The metallic housing (accessi-
ble part) of the device under test is contacted with a test probe or 
alligator clip which is connected to the other input by means of a 
second measurement cable.
Testing the PRO-AB Adapter
The adapter should be tested before use and at regular intervals 
(see adapter operating instructions).
Measuring Sequence
Refer to the operating instructions for the PRO-AB leakage cur-
rent measuring adapter regarding performance of the measure-
ment.
Attention!
!
The test plug should be located in the storage slot during 
leakage current measurement. Under no circumstances 
may the test plug be connected with any system compo-
nents, including PE / ground potential (measured values 
might otherwise be distorted).
The measurement can be started or stopped by pressing the 
“START” key. Leakage current measurement is a long-term mea-
surement, i.e. it continues until it is stopped by the user. The 
momentary measured value is display continuously during mea-
surement.
Note
The self-test must be deactivated in the menu (set 
“TEST ON/OFF” function key to “OFF”) in order to perform a 
measurement.
Always start with the large measuring range (10:1), unless there’s 
no doubt that small measured values can be expected, in which 
case the small measuring range can be used (1:1). The measuring 
range must be selected at the measuring adapter, as well as in 
the menu using the corresponding function key (RANGE). It must 
be assured that the range settings at the adapter and at the test 
instrument are always identical, in order to prevent any distortion 
of measurement results.
Depending on the magnitude of the measured values, the range 
setting can, or must (in the case of overranging), be manually cor-
rected at the measuring adapter and the test instrument.
Individual limit values can be adjusted after pressing the “Limits” 
function key. Exceeded limit values are indicated by the red limit 
value LED at the test instrument.