User Manual
66 GMC-I Messtechnik GmbH
15.2 Creating and Transferring Test Sequences with 
IZYTRONIQ (step-by-step instructions)
➭ Select “Stationary Objects”  .
➭ Then select the “Sequences” menu  .
➭ Click the “Add” symbol   . The “Create New Sequence” field 
is displayed. Enter a “Sequence Name”, a “Test Type” and a 
“Standard”, and select your currently connected instrument 
under “For Device”. Acknowledge by clicking “Add”.
➭ Save the settings by clicking the   icon.
➭ Select the new entry and then the sequence editor   . The 
editing menu appears with “Step Selection” and “Design 
Progress”.
➭ Select the test instrument which is displayed in “Step Selec-
tion”. “Visual Inspection” and “User-Evaluated Measurement” 
appear.
➭ “Test Step: Visual Inspection” is opened in the bottom left-
hand window by dragging “Visual Inspection” into the “Design 
Progress” field. The parameters or details for the respective 
test step must be entered here.
➭ Save the settings by clicking the   icon.
➭ “Test Step: User-Evaluated Measurement” is opened in the 
bottom left-hand window by dragging “User-Evaluated Mea-
surement” into the “Design Progress” field. The parameters or 
details for the respective test step must be entered here.
➭ Save your settings.
➭ Repeat the test steps until the test sequence is finished.
➭ Save your settings by clicking the   icon.
➭ Select “Stationary Objects” again  .
➭ Then select the “Export” function   . The export wizard ap-
pears. 
➭ Select the desired test instrument and insert a checkmark 
next to “Sequences”. Select “Export”. The “Export Sequences 
(max. 10)” menu appears.
➭ Mark the sequences to be exported and click the “Export to 
Test Instrument” icon   . 
During the transfer of the 
test sequences a prog-
ress bargraph is shown 
at the PC screen and 
the right-hand image 
appears on the display 
of the test instrument. 
Successful transfer to 
the test instrument via 
IZYTRONIQ is then indi-
cated at the PC.
Note
All test sequences previously stored at the test instru-
ment are deleted. Only those test sequences transferred 
during the most recent export operation via IZYTRONIQ are 
saved at the test instrument.