Instructions

90 GMC-I Messtechnik GmbH
Characteristic Values PROFITEST MASTER & SECULIFE IP
Reference Conditions
Line voltage 230 V ± 0.1%
Line frequency 50 Hz ± 0.1%
Meas. quantity frequency 45 Hz 65 Hz
Measured qty. waveform Sine (deviation between effective
and rectified value 0.1%)
Line impedance angle cos ϕ =1
Probe resistance 10 Ω
Supply power 12 V ± 0.5 V
Ambient temperature + 23 °C ± 2 K
Relative humidity 40% 60%
Finger contact For testing potential difference
to ground potential
Standing surface insulation
Purely ohmic
Nominal Ranges of Use
Voltage U
N
120 V (108 ... 132 V)
230 V (196 ... 253 V)
400 V (340 ... 440 V)
Frequency f
N
16
2
/
3
Hz (15.4 ... 18 Hz)
50 Hz (49.5 ... 50.5 Hz)
60 Hz (59.4 ... 60.6 Hz)
200 Hz (190 ... 210 Hz)
400 Hz (380 ... 420 Hz)
Overall voltage range U
Y
65 ... 550 V
Overall frequency range 15.4 ... 420 Hz
Waveform Sine
Temperature range 0 °C ... + 40 °C
Supply voltage 8 ... 12 V
Line impedance angle Corresponds to cos ϕ = 1 ... 0.95
Probe resistance < 50 kΩ
Power Supply
Rechargeable batteries 8 each AA 1.5 V,
we recommend eneloop type AA HR6,
2000 mAh (article no. Z502H)
Number of measurements (standard setup with illumination)
– For R
INS
1 measurement – 25 s pause:
approx. 1100 measurements
– For R
LO
Automatic polarity reversal / 1 Ω
(1 measuring cycle) – 25 s pause:
approx. 1000 measurements
Battery test Symbolic display of battery voltage
Battery saver circuit Display illumination can be switched off.
The test instrument is switched off
automatically after the last key opera-
tion. The user can select the desired
on-time.
Safety shutdown If supply voltage is too low, the instru-
ment is switched off, or cannot be
switched on.
Recharging socket Installed rechargeable batteries can be
recharged directly by connecting a
charger to the recharging socket: char-
ger for Z502R
Charging time Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
Overload Capacity
R
INS
1200 V continuous
U
L-PE
, U
L-N
600 V continuous
RCD, R
E
, R
F
440 V continuous
Z
L-PE
, Z
L-N
550 V (Limits the number of measure-
ments and pause duration. If overload
occurs, the instrument is switched off
by means of a thermostatic switch.)
R
LO
Electronic protection prevents switching
on if interference voltage is present.
Fine-wire fuse protection
FF 3.15 A 10 s,
Fuses blow at > 5 A
Electrical Safety
Protection class II per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal voltage 230/400 V (300/500 V)
Test voltage 3.7 kV 50 Hz
Measuring category CAT III 500 V or CAT IV 300 V
Pollution degree 2
Fusing, L and N terminals
1 cartridge fuse-link ea.
FF 3.15/500G 6.3 x 32 mm
Electromagnetic Compatibility (EMC)
Product Standard EN 61326-1:2013
Ambient Conditions
Accuracy 0 to + 40 °C
Operation 5 ... + 50 °C
Storage 20 ... + 60 °C (without batteries)
Relative humidity Max. 75%, no condensation allowed
Elevation Max. 2000 m
Mechanical Design
Display Multiple display with dot matrix
128 x 128 pixels
Dimensions W x L x D: 260 x 330 x 90 mm
Weight approx. 2.7 kg with batteries
Protection Housing: IP 40, test probe: IP 40 per
EN 60529/DIN VDE 0470, part 1
Excerpt from Table on the Meaning of IP Codes
Data Interfaces
Type USB slave for PC connection
Type RS 232 for barcode and RFID scanners
Type Bluetooth
®
for connection to a PC
(MTECH+, MXTRA & SECULIFE IP only)
BAT
Interference
emission
Class
EN 55022 A
Interference
immunity
Test Value Feature
EN 61000-4-2 Contact/atmos. –
4 kV/8 kV
EN 61000-4-3 10 V/m
EN 61000-4-4 Mains conn. – 2 kV
EN 61000-4-5 Mains conn. – 1 kV
EN 61000-4-6 Mains conn. – 3 V
EN 61000-4-11 0.5 period / 100%
IP XY
(1
st
digit X)
Protection Against Foreign
Object Entry
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
4 1.0 mm dia. 0 Not protected