Instructions
GMC-I Messtechnik GmbH 83
Messages — LCD Pictographs — Test sequences
AUTO
The test sequence includes a measurement which cannot be processed 
by the connected test instrument. The respective test step must be 
skipped. Example: The test sequence includes a RCM measurement 
which has been transferred to PROFITEST MTECH.
AUTO The test sequence has been successfully completed.
AUTO
There are no test sequences available.
Cause: They may have been deleted by the following operations: Chang-
ing of language, profile, DB mode or by resetting the test instrument to 
default values.
Error Messages — LCD Pictographs — PRO-AB Leakage Current Adapter
EXTRA → I
L
Measuring range exceeded
Change into the bigger measuring range (test instrument and leakage 
current measuring adapter).
EXTRA → I
L
Test measurement:
The test has been passed sucessfully.
The leakage current measuring adapter is now ready for use.
EXTRA → I
L
Test measurement:
The test has failed.
The leakage current measuring adapter is defective. 
Please consult our repair service.
EXTRA → I
L
Test measurement:
Check the fuse in the leakage current measuring adapter.
Status
Test 
plug
Meas. 
Adapter
Position of the 
Function Switch
Function / Meaning