User Manual
24 GMC-I Messtechnik GmbH
Wiring Diagram 
Measurement of RPE at Single-Phase Extension Cords with EL1
– Measurement type PE(TS) - P1 (passive) 
– Extension cord plug connected to test socket
– EL1 to P1 terminals
Schematic Diagram 
Wiring Diagram 
Protection Class I Devices 
Special Case: Line Voltage at Test Socket (for testing PRCDs) 
– Measurement type PE(TS) - P1 (active) 
– DUT mains plug to test socket 
– Test probe P1 to P1 terminals
Schematic Diagram 
Protective conductor resistance is measured between the earth-
ing contacts at the mains plug and the earthing contact con-
nected to the housing by contacting the housing with test probe 
P1.
Wiring Diagram 
Protection Class I Devices 
Special case: permanently installed DUTs 
– Measurement type PE(mains) - P1 
– Test probe P1 to P1 terminals
Schematic Diagram 
In the case of permanently installed DUTs, protective conductor resis-
tance is measured between the mains power earthing contact 
and the earthing contact connected to the housing by contacting 
the housing with test probe P1.
Wiring Diagram 
Measurement via current clamp sensor 
at permanently installed DUT 
– Measurement type PE(mains) - P1 clamp 
– Test probe P1 to P1 terminals 
– Clamp to COM-V (only withSECUTEST PRO or feautureI01 with 
optional WZ12C current clamp sensor)
Schematic Diagram 
Measurement of test current by closing the current clamp sensor 
around mains PE and contacting the housing with test probe P1 
for permanently installed protection class I devices under test