10.2 Selecting and Configuring a Test Sequence
Sample: Initial Page of a Test Sequence – Tree View
Sample: Initial Page of a Test Sequence – Detail View and
Applied Part
* SETUP switch position:
Setup menu 1/3 > Auto. Measurements > 2/2 > Initial Window Style:
Tree or Detail View
Classification Parameter – Automatic Detection
If the settings for certain classification parameters are automati-
cally detected by the test instrument, this is indicated respectively by
an orange frame (as of firmware V1.3.0; here: test socket connection
type and protection class I). Descriptions of these parameters are
listed in the following tables relative to the respective switch posi-
tions.
Automatic recognition for safety class active
When connecting or disconnecting a DUT, the
safety class may be changed without request for
confirmation where necessary.
Automatic recognition for safety class active
The test instrument retains the selected safety class
setting when a DUT is connected or disconnected.
Classification Parameter – VDE 0701-0702
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
Classification Parameter – VDE 0701-0702-IT
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
A1
Standard/
Connection
type
Protection class
Class parameter
Measurement type
Status line
Sequence parameters
Tree view*
Test sequence
A1
Standard/
Connection Type Protection class
Class parameter
Measurement type
Status line
Sequence parameters
Applied parts
Detail view*
Test sequence
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence
VDE 0701-0702
VDE 0701-0702-IT, see following table
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: VL2E-Adapter
Adapter: AT3-Adapter (feature I01)
Perman. Con.: P1+P2 (only with feature H01)
2/2
Meas. Mode
1)
passive
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter (feature I01)
2/2
Meas. Mode
1
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.