User Manual
GMC-I Messtechnik GmbH 55
Classification Parameter – IEC 62353
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Classification Parameter – IEC 60974-4
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Sequence Parameter (as of firmware V1.3.0)
The default test sequences can be adapted to your applica-
tion or test standard via the sequence parameter. The entered
sequence parameter settings are only valid for the currently
selected switch position (A1 to A8 or AUTO) and are retained until
they are changed. Not all of the parameters are relevant, depend-
ing on the selected DUT classification (protection class etc.).
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see table above
VDE 0701-0702-ExtC, see table above
VDE 0701-0702-PRCD, see previous table
IEC 62353
IEC 60974-4, see following table
Protection class
1
Class I, Class II or Class I+II
Connection type
1
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter
Perman. Con.: P1+P2 (only with feature H01)
2/2
Measurement type (MA)
1
passive
active
Applied parts Applied parts: none, B, BF, CF or combinations
Type B (body): Devices of this type are suitable for both
internal and external patient applications, except for use in
direct proximity to the heart.
The following protection classes are permissible:
I, II, III or devices with internal electrical power supply.
Type BF (body float): same as type B, but with type F in-
sulated applied parts.
Type CF (cardiac float): Devices of this type are suitable
for use directly at the heart. The applied part may not be
grounded.
The following protection classes are permissible:
I, II or devices with internal electrical power supply.
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protec-
tion class must be entered manually.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see table above
VDE 0701-0702-ExtC, see table above
VDE 0701-0702-PRCD, see table above
IEC 62353, see previous table
IEC 60974-4
Protection class
1
Class I, Class II or Class I+II
Connection type
1
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter
2/2
Measurement type (MA)
1
Active
Voltage, rating plate Voltage from rating plate U(R) RMS
(Limit value RMS, variably adjustable)
or open-circuit voltage U0 DC (limit value = 113 V DC)
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protec-
tion class must be entered manually.