User Manual
34 GMC-I Messtechnik GmbH
Test Sequence
Attention!
!
Prerequisite for Testing 
The measurement of insulation resistance may not be 
co
nducted on protection class I devices which have not 
passed the protective conductor resistance test.
Note
The insulation test cannot be performed for all DUTs, for 
example electronic devices, EDP equipment, medical 
devices etc. Leakage current measurements must be 
performed for these DUTs (see 
Section 8.7). 
Observe the notes in the service instructions.
Attention!
!
In order to prevent damage to the instrument, measure-
ment of insulation resistance
 may only be performed be-
tween application parts, measurement inputs or 
int
erfaces and the protective conductor or the housing if 
the instrument is laid out for measurements of this type.
Attention!
!
Touching the DUT During Measurement 
Testing is conducted with up to 500 V, and although cur-
rent is limited (I < 3.5 mA), if the DUT is touched electrical 
shock may occur which could result in consequential ac-
cidents.
Attention!
!
Switch Settings at the DUT 
All switches at the DUT must be set to the on position 
duri
ng measurement of insulation resistance, including 
temperature controlled switches and temperature regula-
tors as well. 
Measurement must be performed in all program steps 
for devices equipped with program controllers.
➭ Set the rotary switch to the R
ISO
 position.
➭ Select the measurement type and the test voltage.
➭ Th
e Up– and Up+ keys provide you with direct access to the 
test voltage parameters: each time this key is pressed, the 
setpoint value shown in the measuring window, Up(set), is re-
duced or increased by 10 V.
➭ Connect the D
UT to the test socket.
➭ S
tart the test: press the START/STOP key.
Note
The measurement is disabled if a voltage of greater than 
25 V is measured between the terminals.
➭
The measured values are displayed. The measured 
value recording symbol shown at the right appears. 
Each time
 this key is pressed, the currently displayed 
measured value is saved to the buffer.
Attention!
!
Removing the Connector Cable 
Do not remove the DUT’s connector cable until the test 
ha
s been stopped, in order to assure that the capacitors 
have been discharged.
➭ Stop the test: press the START/STOP key. 
The save symbol appears (floppy disk showing the 
number of measured values 
stored to buffer memory) 
and prompts you to save the measured values to an ID 
number.
➭ Read the measur
ed values and compare them with the 
table of permissible limit values.
➭ Press the ESC key in order to discard the measured 
values stored to buffer memory and acknowledge by 
pressing the key shown at the right. 
Minimum Allowable Limit Values for Insulation Resistance 
Tes t 
Standard
Test Volta ge
R
ISO
LNPE
LN
Probe
Probe
PE
PC III  Heating
VDE 0701-
0702:200
8
500 V
1 M 2 M 5 M 0.25 M 0.3 M *
DIN EN 60974-4
VDE 0544-
4:2009-06
2 M 5 M 5 M
* With switched on heating elements 
(if heating power > 3.
5 kW and R
ISO
 < 0.3 M: leakage current mea-
surement is required)
Test Standard
Test 
Vo
ltage
R
ISO
PC I PC II
IEC 62353
(VDE 0751-1)
500 V
2 M 7 M
BF or CF BF or CF
70 M 70 M
Notes
Insulation resistance and/or leakage current must be measured 
by contacting all exposed, conductive parts with test probe P1 for 
protection class II and III devices, as well as for battery powered 
devices.
Batteries must be disconnected during t
esting of battery powered 
devices.
8.7 Measuring Leakage Current
Attention!
!
Measurement with DUT Connected to Line Voltage 
It’s absolutely essential to assur
e that the device under 
test is operated with line voltage during performance of 
leakage current measurements with the direct or differential 
current method. Exposed conductive parts may conduct 
dangerous touch voltage during testing, and may not 
under any circumstances be touched. (Mains power is 
disconnected if leakage current exceeds approx. 
10 mA.)
Frequency response in 
acc
ordance with the 
figure to the right is 
taken into consider-
ation for all leakage 
curr
ent measure-
ments (IPE, IB, IG, IP) 
(dir
ect, differential, 
alternative). 
10 10
2
10
3
10
4
10
5
10
6
+20
0
–20
–40
–60
Frequency (f) in Hz
Relative Magnitude (dB):
U(f)
U(f=10)










