Alternative Measuring Method (equivalent leakage current) 
– Alternativ
e Measurement Type (P1) 
– DUT mains plug connected to the test socket 
– Test probe P1 to P1 terminals
Schematic Diagram, Protection Class I
 <help_ig_alt_p1>
After activating test voltage, leakage current is measured between 
short-circuited mains conductors L and N (DUT mains plug) and 
accessible conductive parts (probe contact) which are not con-
nected to the housing. 
If the DUT includes terminals for applied 
parts, they must be 
short-circuited and contacted with test probe P1 as well. 
Wiring Diagram, Protection Class I 
Differential Current Measurement 
– AT3-Adapter measurement type 
– DUT mains plug to AT3-I I I E test adapter
– Test probe P1 to P1 terminals
– AT3-I I I E probe to COM-V terminals 
– AT3-I I I E test plug to test socket
Schematic Diagram 
Measurement at the DUT with 3-phase mains connection via 
AT3-I I I E adapter
Wiring Diagram  
Measurement Method with Current Clamp Sensor 
for Permanently Installed DUTs 
– Clamp m
easurement type 
– Clamp to COM-V (only withSECUTEST PRO or feautureI01 
 with optional WZ12C current clamp sensor)
Schematic Diagram 
Measurement of device leakage current by closing the current 
clamp sensor around the L and N conductors in the power cable 
of the mains for permanently installed protection class I devices 
under test
Wiring Diagram  
Set Measuring Range at Current Clamp Sensor and Parameter at 
the
 SECUTEST PRO
SECUTEST PRO
 Current Clamp Sensor SECUTEST PRO
Transformer Ratio 
Pa
rameter
Transformer Ratio 
(Switch*)
Measuring 
Range
Display Range 
with Clamp
1 mV : 1 mA
WZ12C
1 mV : 1 mA 1 mA... 15
 A 0 mA ... 300 A
100 mV : 1 mA **
SECUTEST CLIP
100 mV : 1 mA 0.1...25 mA 0.00 mA ... 3.00 A
* only with WZ12C
** Default value