48 GMC-I Messtechnik GmbH
8.8 Probe Voltage – U
U
Single measurements, rotary switch level: green
Switch
Position
Measure-
ment Type,
With Mains 
to Test 
Socket
Measure-
ment Type,
Without 
Mains 
to Test 
Socket
Measuring Functions
U
PE - P1
U
  Probe voltage, RMS
U
~ 
Alt. voltage component
U
= 
Direct voltage component
PE - P1 
(with mains)
U  Probe voltage, RMS
U
~ 
Alt. voltage component
U
= 
Direct voltage component
Mains to Test Socket
Schematic Diagram  
Wiring Diagram
Permanently Connected DUT
Schematic Diagram 
Wiring Diagram
Direct, alternating and pulsating voltages of up to 253 V can be 
measured. Two connection types are available, one of which has 
to be selected in the parameters menu. 
Setting Measuring Parameters for U
Probe
Measuring 
Parameter
Meaning
Measurement Type,
Suitable for 
DUT Connection via
PE-P1 Measurement of voltages with 
r
eference to PE, test socket re-
mains voltage-free
Permanent connection
PE-P1 (with 
ma
ins)
Measurement of voltages with 
reference to PE, line voltage is 
applied to the test socket 
Test socket
Polarity – only for PE-P1 (with mains)
L/N or N/L Selection of polarity for mains voltage to the test socket
Test Sequence
➭ Set the rotary switch to the U position.
➭ Connect the DUT’s mains plug to the test instrument’s test 
socket.
➭ Start the test: press the START/STOP key. 
➭ PE-P1 (with mains
): Acknowledge the warning which in-
dicates that line voltage will 
be connected to the test 
socket. 
➭ Co
ntact the ungrounded output for safety extra-low voltage 
with test probe P1.
➭
Polarity can be set via direct selection immediately be-
fore measurement is started, without having to switch to 
th
e parameters menu.
➭
The measured values are displayed. The measured 
value recording symbol shown at the right appears. 
Each time th
is key is pressed, the currently displayed 
measured value is saved to the buffer.
➭ S
top the test: press the START/STOP key. 
The save symbol appears (floppy disk showing the 
number of measured values 
stored to buffer memory) 
and prompts you to save the measured values to an ID 
number.
➭ P
ress the ESC key in order to discard the measured 
values stored to buffer memory and acknowledge by 
pressing the key shown at the right. 