Set measurement duration of individual test steps (as of firmware 1.5.0)
Testing time for the respective measurement can be influenced 
with these parameters. If a test step for a single measurement is 
involved, the entire test step has a duration of the time entered in 
seconds. If a test step for a multiple measurement is involved, the 
measurement duration for each measuring point is influenced.
If 0 seconds is selected, continuous measurement is conducted 
which can 
only be ended by pressing a key.
Parameter 
(as of FW 1.5.0)
Meaning
RPE measurement 
duration 
1
Set testing time for the protective conductor resistance 
measurement (0 to 60 seconds)
IPE measurement 
duration
Set testing time for the protecti
ve conductor current 
measurement (0 to 60 seconds)
IG measurement 
duration
Set testing time for the device l
eakage current measure-
ment (0 to 60 seconds)
1
For measuring sequence VDE 0701-0702-PRCD with the following set-
ting „PRCD type: PRCD (SPE)“ th
e measurement duration cannot be in-
fluenced.
The measurement duration which has been se
t in this case only affects 
the RPE measurement with PRCD types „PRCD (standard)“ and 
„PRCD-S (SPE)“.
Parameter 
(as of FW1.8.0)
Meaning
IB measurement 
duration
Set testing time for the touch current measurement
(0 to 60 seconds)
IB SK II measurement 
duration
(only for IEC 60974) Set testing time for the 
touch current 
measurement at PC II parts (with the exception of welding 
outputs) (0 to 60 seconds)
RISO SK II measurement 
duration
Set testing time for RISO measurements at PC II parts 
(0 to 60 seconds)