Protection Class I Devices 
Special Case: Line Voltage at Test Socket (for testing PRCDs) 
– Measurement type PE(TS) - P1 (active) 
– DUT mains plug to test socket 
– Test probe P1 to P1 terminals
Schematic Diagram 
Protective conductor resistance is measured between the earth-
ing contacts at the mains plug and the earthing contac
t con-
nected to the housing by contacting th
e housing with test probe 
P1.
Wiring Diagram  
Protection Class I Devices 
Special case: permanently installed DUTs 
– M
easurement type PE(mains) - P1 
– Test probe P1 to P1 terminals
Schematic Diagram 
In the case of permanently installed DUTs, protective conductor resis-
tance is measured between the m
ains power earthing contact 
and the earthing contact connected to the housing by contacting 
the housing with test probe P1.
Wiring Diagram  
Measurement via current clamp sensor 
at permanently installed DUT 
– M
easurement type PE(mains) - P1 clamp 
– Test probe P1 to P1 terminals 
– Clamp to COM-V (only withSECUTEST PRO or feautureI01 with 
optional WZ
12C current clamp sensor)
Schematic Diagram 
Measurement of test current by closing the current clamp sensor 
around mains PE and contacting the housing with test probe P1 
for permanently installed protection class I devices under test
Wiring Diagram  
Set Measuring Range at Current Clamp Sensor and Parameter at 
th
e SECUTEST PRO or 
SECULIFE ST BASE
This measurement type can only be selected if test current is set 
to 10 A AC. 
SECUTEST PRO  Current Clamp Sensor SECUTEST PRO
Transformer Ratio 
Pa
rameter
Transformer Ra-
tio 
(
Switch*)
Measuring 
Range
Display Range 
with Clamp
1 mV : 1 mA
WZ12C
1 mV / mA 1 mA... 15 A 0 mA ... 300 A
100 mV:1 mA
SECUTEST CLIP
100 mV/mA
0.1...25 mA 0 mA ... 300 A
* only with WZ12C
2-Pole Measurement at Permanently Installed DUTs 
(only with SECUTEST PRO or feature H01) 
– M
easurement type P1 - P2 
– Test probe P1 to P1 terminals
– Test probe P2 to P2 terminals
Schematic Diagram 
PE at the mains connection is contacted with the second test 
probe instead of via the test instrument’s mains plug.