11.1 List of error messages
Error Messages Possible Causes Corrective Measures
Mains Connection Errors
– Protective conductor PE at the 
mains outlet at which the SECUT-
EST is being operated is conduct-
ing voltage! This de
tection function 
makes use of the metallized 
START/STOP key on the test 
instrument. In order for detection to 
function correctly, it must be possi-
ble to establish reference to earth 
pot
ential via the user’s finger.
Note
If the user’s finger is insulated 
from the key when it’s 
pressed, this error message 
may occur although the instal
-
lation is OK (see “Automatic 
Recognition of Mains Con-
nection Errors” on page 10).
➭ Please
 remove the SECUTEST’s 
mains plug from this outlet and ar-
range to have the outlet/installa-
tion inspected by a qualified 
electr
ician without delay. Do not 
operate any other devices at this 
electrical outlet before this inspec-
tion has been completed.
➭ In 
order to ensure that detection 
functions reliably, repeat the inter-
ference voltage test and observe 
the following tips: 
– Unplug all USB devices from the 
S
ECUTEST’s USB ports.
– Remain in contact with a 
gr
ounded object while pressing 
the START/STOP key (e.g. a 
heating pipe).
– Do not contact the START/STOP 
key with an object or while wearing 
gloves.
PE connection not detected (at the 
outlet at which the test instrument is 
being operated):
– If the installation is defective!
– In the case of special types of TT 
syst
ems; detection may fail in this 
case.
– If the test instrument is being oper-
ated in an IT system
➭ If the
 test instrument is being oper-
ated in an IT system: 
Acknowledge the question by 
pressing 
 – the IT system option 
is activated in this case.
➭ If i
t’s not an IT system: remove the 
mains plug from the outlet and in-
spect the installation without delay!
➭ If it
’s a TT system without neutral 
conductor, press 
; direct leak-
age current measurements are 
possible. 
(Make absolutely sure that direct 
lea
kage current measurements are 
possible in your current mains 
type!)
As opposed to the previously used 
mains connection, PE was detected 
while the IT system option was acti-
vated in setup.
➭ Operation in an IT system: Re-
spond to the question by pressing 
. 
The IT System option remains thus 
active.
➭ Oper
ation in an TN or a TT system: 
Respond to the question by press-
ing 
. As a consequence, the IT 
system option is deactivated.
Line frequency is less than 48 or 
greater than 62 Hz.
➭ PE detection does not work in this 
case: select 
 or  , depending 
on whether or not the utilized sys-
tem is an IT system.