User Manual
70 GMC-I Messtechnik GmbH
Database Processing Error
– One of the fields was filled in with 
invalid content while processing an 
existing database object.
➭ Please
 be certain to complete all 
mandatory fields (identified in red).
➭ If necessary, check your entries to 
the fields for invalid special charac-
ters.
– The ID field was not filled in while 
creating a new device.
➭ Fill in the ID field.
– There’s already an object with the 
same ID under the “Customer” 
database object.
An incorrect barcode has been 
sel
ected.
➭ Assign another ID.
Error while creating „.secu“ file on USB 
flash drive
There is not or no longer enough free 
spac
e on storage device left.
– Particularly in the case of USB flash 
dri
ves formatted for FAT16: 
too many files on USB drive
– Power consumption of USB flash 
drive 
in use exceeds 500 mA.
– USB flash drive has been discon-
nected during the data import pro-
cess
– USB flash drive is defective or 
incompatibl
e with the SECUTEST.
Make sure that a minimum storage 
capacity of 100 MB 
is available on the 
USB drive and/or delete any data files 
no longer required.
➭
If the problem persists, save the data 
of the USB drive on another storage 
device and format the USB drive (file 
system FAT32).
➭ Only use USB drives with a power 
consumption of less than 500 mA 
in combination with the SECUTEST
➭ Mak
e sure that the USB drive is not 
disconnected or moved until the 
data export process is completed.
➭ If none 
of these measures leads to 
an improvement, replace the USB 
flash drive. A list of tested devices 
is included in section 14.3.
Error Messages Possible Causes Corrective Measures