User Manual
GMC-I Messtechnik GmbH 77
1
Specified values are only valid for the display at the test instrument. Data 
transmitted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current 
from previous standards
3
Protective conductor current, touch current, device leakage current, pa-
tient leakage current
4
Protective conductor current, touch current, device leakage current
5
Only with 
SECUTEST BASE10
 (feature G01), 
SECUTEST PRO
 or 
SECULIFE ST BASE
6
Only with SECUTEST PRO (feature I01) or SECULIFE ST BASE
7
Measurement types IPE_clamp and IG_clamp
8
Measurement type IPE_AT3 adapter and IG_AT3 adapter
9
The upper range limit depends on the selected test voltage.
10
Due to inrush current limiting components, the voltage at the test socket 
may be lower than the measured line voltage.
11
SECULIFE ST BASE 25 only
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
Test times (“measurement duration” parameter) can be set separately 
for each rotary switch position during configuration of the sequence 
parameters. Test times are neither tested nor calibrated.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA), 
automatic shutdown ensues within 100 ms. This shutdown does 
not tak
e place during leakage current measurement with current 
clamp sensor or adapter.
Influencing Quantities and Influence Error 
Influencing Quantity /
Sphere of Influence
Designation
per 
IEC 61557-16
Influence error
 % r
dg.
Change of position E1 —
Change to test equipment supply 
vol
tage
E2 2.5
Temperature fluctuation
E3
Specified influence error valid 
start
ing with temperature changes 
as of 10 K:
0 ... 40 
C 2.5
Amount of current at DUT E4 2.5
Low frequency magnetic fields E5 2.5
DUT impedance E6 2.5
Capacitance during insulation 
measurement
E7 2.5
Waveform of measured current
E8
49  51 Hz
2 with capacitive load (for equiva-
lent leakage current)
45  100 Hz 1 (for touch current)
2.5 for all other measuring ranges
I
Clamp
Current via 
current clamp 
sensor 
1 mV : 1 mA] 
(V–COM sockets
6 7
)
1 ... 99 mA
1 mA
(1 mV)
— — — — — — —
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z 
without clamp
253 V Cont.
0.1 ... 0.99 A
0.01 A
(10 mV)
1.0 ... 9.9 A
0.1 A
(100 mV)
10 ... 300 A
1 A
 (1 V)
Current via 
current clamp 
sensor 
[10 mV : 1 mA] 
(V–COM sockets
6 7
)
0.1 ... 9.9 mA
0.1 mA
(1 mV)
— — — — — — —
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z 
without clamp
253 V Cont.
10 ... 99 mA
1 mA
(10 mV)
0.10 ... 0.99 A
0.01 A
(100 mV)
1.0 ... 30.0 A
0.1 A
(1 V)
Current via 
current clamp 
sensor 
[100 mV : 1 mA] 
(V–COM sockets
6 7
)
0.01 ... 0.99 mA
0.01 mA
(1 mV)
— — — — — — —
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z 
without clamp
253 V Cont.
1.0 ... 9.9 mA
0.1 mA
(10 mV)
10 ... 99 mA
1 mA
(100 mV)
0.10 ... 3.00 A
0.01 A
(1 V)
Current via 
current clamp 
sensor 
[1000 mV : 1 mA] 
(V–COM sockets
6 7
)
1 ... 99 μA
1 μA
(1 mV)
— — — — — — —
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z 
without clamp
253 V Cont.
0.10 ... 0.99 mA
0.01 mA
(10 mV)
1.0 ... 9.9 mA
0.1 mA
(100 mV)
10 ... 300 mA
1 mA
(1 V)
Func-
tion
Measured 
Quantity
Display Range /
Nominal Range of 
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit 
Voltage
U
0
Nom. 
Current
I
N
Short-
Circuit
Current
I
K
Int.
Resis-
tance
R
I
Ref. 
Resis-
tance
R
REF
Measuring 
Uncertainty 
1
Intrinsic Error 
1
Overload 
Capacity
Value Time