The voltage measuring range is set manually to 10 V.
The first measured value is 5 V and is stored to memory because
it is greater than 10% of the measuring range. As soon as the
measured value drops to less than 10% of the measuring range,
i.e. amounts to less than 1 V which corresponds to removal of the
test probes from the measuring point, the instrument is ready to
store a new value.
4.4.1 Saving Minimum and Maximum Values – Min/Max Function
Minimum and maximum measured values applied to the
measuring instrument’s input after the Min/Max function has been
activated can be “frozen” at the display. The most important use
of this function is the determination of minimum and maximum
values during long-term measured value observation.
The Min/Max function can be activated in all measuring functions.
Apply the measured quantity to the instrument and set the mea-
suring range with the MAN / AUTO key before activating the Min/
Max function.
The Min/Max function is deactivated by pressing and holding the
DATA/MIN/MAX key (approx. 1 second), when the measuring func-
tion is changed or when the instrument is switched off and back
on again.
Note
As opposed to the DATA function, the Min/Max function can
also be used for temperature measurement.
Min-Max
Function
Data/
Min/Max
Key
Min. and Max.
Measured Values
Response from Instrument
Display
Acous
tic
Sig-
nal
Digital
Measured
Value
Max.
Min.
1
Activate
and save
1 x brief Are saved
Current
measured
value
Min.
Once
2
Save and
display
Brief
Storage continues in back-
ground, new min. and max.
values are displayed together
with the point in time of
occurrence.
Saved min.
value
Min.
Once
Brief
Saved max.
value
Max.
Once
Stop Long Are deleted
Current
measured
value
Is
deleted
Twice