V/Hz, Ω, Temperature, and A/Hz Measurements
5.5 Diode Testing with Constant Current of 1 mA
Ð Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
Ð Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to section 5.2.1 regarding testing for the absence of volt-
age with the help of the direct voltage measurement.
Ð Set the rotary switch to .
Ð Connect the device under test as shown.
Attention!
!
Observe open-circuit voltage of 7 V during diode testing.
Circuits must be laid out accordingly.
Forward Direction and Short-Circuit
The instrument displays forward voltage in volts (display:
4 places). As long as voltage drop does not exceed the maximum
display value of 6 V, several series connected components or
reference diodes can be tested with a small reference voltage and
Z-diodes.
Reverse Direction and Interruption
The measuring instrument indicates overload 0L.
Note
Resistors and semiconductor paths connected in parallel to
the diode distort measurement results!
–+ +
Forward Direction Reverse Direction
0V !
0.654
0.L
to 6 V
Measuring range:
+
–
!
1.8569
V
DC