V/Hz, Ω, Temperature, and A/Hz Measurements
5.8.4 AC Measurement with Current Clamp Sensor, A AC and Hz
Voltage/Current Transformer Output
When a current clamp sensor is connected to the multimeter
( V input / mA input) all current displays appear with the cor-
rect value in accordance with the selected transformation ratio.
The only prerequisite is that the current sensor is equipped with at
least one of the below listed transformation ratios, and that the
ratio has been previously selected in the following menu
(Cl ip 0FF) (see also section 7.4).
Current Clamp Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff

The maximum allowable operating voltage is equal to the nominal
voltage of the current transformer.
When reading the measured
value, additional error resulting from the current clamp sensor must
also be taken into consideration.
(default setting: OFF)
Transforma-
tion Ratios
CL IP
DMM Measuring Ranges Clamp Types
300 mV 3 V 30 V
1:1
1mV/1mA
300,00 mA 3,0000 A 30,000 A WZ12C
1:10
1mV/10mA
3,0000 A 30,000 A 300,00 A WZ12B, Z201A
1:100
1mV/100mA
30,000 A 300,00 A 3000,0 A Z202A
1:1000
1 mV/1 A
300,00 A 3000,0 A (30000,0 A)
Z202A, Z203A,
WZ12C
MEAS / CAL
SETUP
003.50
A
A~
Hz
050.10
Hz
+ +
~
R
x
Current
R
i
= 1 MΩ/9 MΩ
mA~
ZERO/SEL
ESC
AC