User Manual

60 GMC-I Messtechnik GmbH
Device and Measuring Parameters
7 Device and Measuring Parameters
The instrument’s SETmode (menu mode) makes it possible to
set operating and measuring parameters, query information and
activate the interface.
Ð The menu mode is accessed by pressing the
MEAS / CAL | SETUP key, assuming that the instrument is
switched on and set to “Measure” (measuring mode).
1nFo” appears at the display.
Ð The main menus, i.e. the “SETUP”, tEMP” “SEnd” and “StorE
menus, are accessed, and the display is once again returned to
InFo, by activating the
 keys (in any direction).
Ð After selecting the desired main menu, sub-menus are
accessed by pressing the OUT | ENTER key.
Ð The desired parameter is selected by repeatedly pressing the
 key.
Ð In order to check or change a parameter, acknowledge it with
the OUT | ENTER key.
Ð The
keys can be used to position the cursor at the entry
position. The desired value is selected with the help of the 
keys.
Ð Changes can only be accepted with the OUT | ENTER key.
Ð You can return to the sub-menu without making any changes
by pressing the ZERO / SEL | ESC key, and to the main menu by
pressing the ZERO / SEL | ESC key once again etc.
Ð You can switch to the measuring mode from any menu level by
pressing the OUT | ENTER key.
After repeatedly pressing the MEAS / CAL | SETUP key (without first
turning the multimeter off), you can return to the last selected
menu or parameter from the measuring mode.
Example: Setting Time
1nFo ... set rAte ... t iME
10:24 (hh:mm)

Setting hours and minutes:
Advance to desired entry position.
 Change the setting, the entry position blinks.
Press and hold the key
to change the setting rapidly.
The new time setting is activated after
acknowledgment.
MEAS / CAL
SETUP