User Manual

GMC-I Messtechnik GmbH 63
Device and Measuring Parameters
APoFF – specified time for automatic shutdown and continuous ON
DMM
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time, and if none of the
keys or the rotary switch have been activated before the specified
time “APoFF” (entered in minutes) has elapsed.
Calibrator
The instrument switches the output off automatically if none of the
keys or the rotary switch have been activated during the time
specified in minutes in “APoFF. The display is also switched off
after one additional minute.
Setting APoFF
If the on setting is selected, the multimeter is set to continuous
operation and
ON appears in the display to the right of the battery
symbol. In this case, the multimeter can only be switched off
manually. The “on” setting can only be canceled by changing the
respective parameter, and not by switching the instrument off.
1nFo
... SET time ... APoFF
10 ... 59 min on

(10 minutes = default setting)
bEEP – set limit value for continuity testing
1nFo
... SET time ... bEEP
1, 10, 20, 30, 40 ... 300 Ω 
(10 Ω = default setting)
irStb – status of infrared receiver in the stand-by mode
See section 8.2 on page 67 regarding settings.
Addr – set device address
See section 8.2 on page 67.