User Manual

70 GMC-I Messtechnik GmbH
Technical Data
1)
Display: 3¾ places for capacitance measurement; a different resolution and
sampling rate can be selected in the rAtE menu for saving and transmitting
measured values.
3)
At 0 ° ... + 40 °C
6)
Lowest measurable frequency for sinusoidal measuring signals symmetrical to the
zero point
7)
Range 60/300 mV~: U
E
30% of upper range limit
3/30/300 V~: U
E
10% of upper range limit
For voltage > 100 V: power limiting: 6 · 10
6
V · Hz
8)
Plus sensor deviation
9)
Without integrated reference junction;
with internal reference temperature plus error ±2K
10)
The limits only apply for battery operation (mains adapter Z218K for multimeter
operation in preparation)
Key
d = digit MR = measuring range rdg. = reading (measured value)
Meas.
Func-
tion
Measuring Range
Resolution at Upper
Range Limit
Input Impedance
Intrinsic Uncertainty at Highest Resolution
under Reference Conditions
Overload Capacity
3)
Value Time
30,000
(60,000)
1)
3100
1)
f
min
6)
±(...% rdg. + ... d)
Hz
300 Hz 0.01 Hz
1Hz
0.05 + 5
7) 10)
300 V
Continuous
3kHz 0.1Hz 300 V
30 kHz 1 Hz 200 V
300 kHz 10 Hz 10 Hz 20 V
Meas.
Func-
tion
Temperature
Sensor
Measuring Range
Resolution
Intrinsic Uncertainty at
Highest Resolution under
Reference Conditions
±(...% rdg. + ... d)
8)
OL
Capa-
city
3)
Value Time
°C/°F
Pt100
200.0 100.0 °C
0.1 K
0.3 + 10
300
V
DC
RMS
sine
5
min
100.0 +100.0 °C
+100.0 +850.0 °C
Pt1000
200.0 +100.0 °C
+100.0 +850.0 °C
Ni 100
–60.0 +180.0 °C
Ni 1000
60.0 +180.0 °C
K (NiCr-Ni)
–250.0 +1372.0 °C
0.2 + 10
9)
J (Fe-CuNi)
–210.0 +1200.0 °C
T (Cu-CuNi)
–270.0 +400.0 °C
B (Pt30Rh/
Pt6Rh)
+0 +1820.0 °C
E (NiCr/CuNi)
–270.0 +1000.0 °C
R (Pt13Rh/Pt)
–50.0 +1768.0 °C
N (CU/Cu10)
–270.0 +1300.0 °C
S (Pt10Rh/Pt)
–50.0 +1768.0 °C
J (Fe/CuNi)
–200.0 +900.0 °C
U (Cu/CuNi)
–200.0 +600.0 °C