10 Maintenance
10.1 Batteries
Warning!
Ex Environment: Remove the instrument from the
potentially explosive environment before opening.
Attention!
!
Non-Ex Environment: Disconnect the instrument
from the measuring circuit before opening to re-
place batteries!
Removing the Batteries During Periods of Non-Use
The integrated quartz clock draws power from the batter-
ies, even when the instrument is switched off. It is advis-
able to remove the batteries during long periods of non-
use for this reason (e.g. vacation). This prevents excessive
depletion of the batteries, which may result in damage
under unfavorable conditions.
Checking Battery Voltage-Level and Condition
The current battery voltage level can be queried in the
“Info” menu (see chapter 8 on page 17):
MENU/ESC + ON|OFF  InF0 X.X V (bAtt).
Make sure that no battery leakage has occurred before ini-
tial start-up, and after long periods of storage. Continue to
inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely
clean electrolyte from the instrument with a damp
cloth, and replace the batteries before using the instru-
ment.
If the “ ” symbol appears at the display, the batteries
should be replaced as soon as possible. You can con-
tinue working with the instrument, but reduced mea-
suring accuracy may result.