Datasheet
METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO⏐BASE
TRMS Digital Multimeters
2 GMC-I Gossen-Metrawatt GmbH
Automatic or Manual Measuring Range Selection
Measured quantities are selected by means of a rotary switch and 
a function key. The measuring range is automatically matched to 
the measured values. The measuring range can also be selected 
and fixed manually with a key.
Fast Acoustic Continuity Test
Testing for short circuiting and interruption is possible with the 
selector switch in the   position. The threshold value for 
acoustic signaling can be set to 1, 10, 20, 30, 40 or 90 Ω.
Automatic Storage of Measured Values *
The DATA function automatically saves the digitally displayed 
measured value after settling in. Acoustic signaling is also used to 
indicate whether the new measured value deviates from the initial 
reference value by less or more than 0.1% of the measuring 
range.
* Patented
Storage of Min-Max Values
Comparable to the slave-pointer function of an analog instrument, 
the device saves the highest and lowest measured values after 
the MIN/MAX function has been activated or reset. These 
extreme values can be queried at the display.
Battery Charging Status – Power Saving Circuit
The battery charging status is indicated by means of four symbols.
The device is switched off automatically if the measured value 
remains unchanged for a period of between 10 and 59 minutes 
(adjustable), and if none of the controls are activated during this time. 
Automatic shutdown can be deactivated by switching the 
instrument to continuous operation.
METRAHIT⏐X-TRA⏐OUTDOOR: The infrared interface can be switched 
off in the standby mode.
Protective Cover for Harsh Conditions
The instrument is protected against damage in the event of impacts 
or dropping by means of a soft rubber cover with tilt stand and test 
probe holder. The rubber material also assures that the instrument 
does not wander if it is set up on a vibrating 
surface.
Infrared Data Interface with METRAHIT⏐X-TRA⏐OUTDOOR
The device can be remote configured, and momentary and stored 
measurement data can be read out via the bidirectional infrared 
interface. The USB⏐X-TRA interface adapter and METRAwin 10 soft-
ware are required to this end (see accessories). Interface 
protocol and device driver software for LabVIEW
®
 (National 
Instruments™) are available upon request.
DKD Calibration Certificate
The multimeters are furnished with an internationally valid DKD 
calibration certificate (recognized by EA and ILAC). A
fter the 
specified calibration interval has elapsed (recommended interval: 
1 to 3 years), the multimeters can be inexpensively recalibrated in 
our own DKD calibration laboratory.
Applicable Regulations and Standards 
Selection List 
1)
For 15,400 measured values, sampling rate adjustable from 0.1 second to 9 hours
Included
1multimeter
1
pair of safety measurement cables (1.5 m) with 4 mm test 
probes,
 1000 V CAT III, 600 V CAT IV (KS17-2)
2 batteries, 1.5 V, type AA
1 condensed operating instructions, English/German
1 CD ROM, content: operating instructions in English and 
German), METRA⏐V
IEW demo software
1 DKD calibration certificate
1 protective rubber cover (METRAHIT⏐X-
TRA⏐OUTDOOR only)
Voluntary Manufacturer’s Guarantee
36 months for materials and workmanship
1 to 3 years for calibration (depending upon application) 
IEC/EN 61 010, part 
1:2001/VDE 0411-1:2002
Safety requirements for electrical equipment for 
measurement, control and laboratory use
DIN EN 61326
VDE 0843, part 20
Electrical equipment for control technology and laboratory use 
– EMC requirements
DIN EN 60529
DIN VDE 0470, part 1
Test instruments and test procedures
– degrees of protection provided by enclosures (IP code)
Function METRAHIT⏐
X-TRA / 
O
UTDOOR
METRAHIT⏐
TECH
METRAHIT⏐
PRO
METRA
HIT
⏐
BASE
V AC / Hz TRMS
(Ri ≥ 9 MΩ) 
•
& Filter
•
& Filter
•
& Filter
•
V AC TRMS
(Ri = 1 MΩ) 
•
& Filter
•
& Filter
•
& Filter
—
V AC+DC TRMS 
(Ri ≥ 9 MΩ) 
••••
V DC (Ri ≥ 9 MΩ)
••••
... 1 MHz 5 V AC
• ———
Keying ratio as %
• ———
Hz (V AC)
... 100 kHz ... 100 kHz ... 100 kHz ... 100 kHz
Bandwidth, V AC
15 Hz ... 20 kHz 15 Hz ... 10 kHz 15 Hz ... 10 kHz 15 Hz ... 1 kHz
A AC / Hz TRMS
100 μA 
1/10/100 mA 
1 A / 10 (16) A
10/100 mA 
1 A / 10 (16) A
1 A / 10 (16) A
—
A AC+DC TRMS
—
A DC
—
Fuse
10 A/1000 V 10 A/1000 V 10 A/1000 V —
Transformation factor 
— • — •
A AC  / Hz TRMS
—
mV/A
mA/A
—
mV/A 
Ri = 1 MΩ
A AC+DC   TRM
S
—
mV/A
mA/A
—
mV/A 
Ri = 1 MΩ
A DC 
—
mV/A
mA/A
—
mV/A 
Ri = 1 MΩ
Hz (A AC)
... 30 kHz ... 30 kHz ... 30 kHz ... 30 kHz
Resistance  Ω
••••
Continuity 
••••
Diode ... 5,1 V
••••
Temperature TC (K)
••••
Temperature RTD
• ———
Capacitance 
••——
MIN/MAX / data hold
••••
4 MBit memory 
1)
• ———
IR Interface
• ———
Power pack adapter 
socket
• ———
Protection
IP52 / IP65 IP52  IP52  IP52
Measuring category
1000 V CAT III
600 V CAT IV
1000 V CAT III
600 V CAT IV
1000 V CAT I II
600 V CAT IV
1000 V CAT III
600 V CAT IV
1kHz 1kHz 1kHz