Operation Manual
42 GMC-I Messtechnik GmbH
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.7.2 Alternating Current and Frequency Measurement, 
Direct Connection, A AC and Hz 
(METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only)
Ð First disconnect supply power from the measuring circuit or 
the power consumer (1), and discharge any capacitors.
Ð In accordance with the current or frequency to be measured, 
turn the rotary switch to A~ or Hz.
Ð Select the desired measured quantity by briefly pressing the 
FUNC | ENTER multifunction key. Each time the key is pressed, 
AC
TRMS
 and Hz are alternately selected, and 
switching is acknowledged with an acoustic signal.
Ð Safely connect the measuring instrument (without contact 
resistance) in series to the power consumer as shown.
Ð Switch supply power to the measuring circuit back on (3).
Ð Read the display. Make a note of the measured value if the 
instrument is not being operated in the memory mode or the 
transmission mode.
Ð Disconnect supply power from the measuring circuit or the 
power consumer (1) once again, and discharge any 
capacitors.
Ð Remove the test probes from the measuring point and return 
the measuring circuit to its normal condition.
COM mA A
V
Temp
Ω
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
003.50
A
AC
TRMS
050.10
Hz
AC
8
.
8
.
8
.
8
.
8
!
I > 10 A
!
I > 16 A
Current measurement 
may only be performed 
with installed batteries!
Measuring Ranges:
Current
FUNC
ENTER
METRAHIT⏐X-TRA/OUTDOOR
METRAHIT⏐PRO
1 A / 10 A (16 A max. 30 s)
100 μA / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRAHIT⏐T
ECH
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)