Operation Manual
GMC-I Messtechnik GmbH 43
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.7.3 Direct and Pulsating Current Measurement with Clip-On Current Sen-
sor, A DC and A (DC+AC) (METRA
HIT
⏐
T
ECH
 and METRA
HIT
⏐
B
ASE
 only)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter 
(METRAHIT⏐T
ECH: V input, METRAHIT⏐BASE:  V input), all current 
displays appear with the correct value in accordance with the se-
lected transformation factor.
The only prerequisite is that the current 
sensor is equipped with at least one of the below listed transformation 
factors, and that the factor has been previously selected in the following 
menu (
Cl
ip
≠ 
0FF
), see also chapter 6.4.
Current Clip Setup Menu
 1nFo 
 ...  SET   rAtE  ...  CLIP 
 1 / 10/100/1000 / 0ff  
Maximum allowable operating voltage is equal to the current 
transformer’s nominal voltage. When reading the measured value, 
additional error resulting from the clip-on current sensor must also 
be taken into consideration.
(default setting: 
METRAHIT⏐TECH: OFF, METRAHIT⏐BASE: 1:1000)
Trans. Factors
CL IP
Measuring Ranges DMM Clip Types
100 mV 1 V 10 V
1:1 
1 mV / 1 mA
100.00 mA 1.0000 A 10.000 A WZ12C
1:10 
1m V / 10 mA
1.0000 A 10.000 A 100.00 A
WZ12B, Z201A/B, 
METRAFLEX
1:100 
1m V / 100 mA
10.000 A 100.00 A 1000.0 A
Z202A/B, 
METRAFLEX
1:1000 
1 mV / 1 A
100.00 A 1000.0 A (10000.0 A)
Z202A/B, Z203A/B,
WZ12C, 
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM
V
Temp
Ω
A
~
R
x
Current
003
.
50
A
1:1000
A
DC
003
.
50
A
1:1000
A
DC
AC
TRMS
A
A
FUNC
ENTER
R
i
R
i
 = 9 M
Ω
METRAHIT⏐TECH
COM
V
Temp
Ω
V
~
R
x
Hz
Current
R
i
R
i
 ~ 1 M
Ω
METRAHIT⏐BASE
mA
METRAHIT⏐BASE
METRAHIT⏐TECH
V
V










