Operation Manual

GMC-I Messtechnik GmbH 49
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
5.9.3 Direct and Pulsating Current Measurement
with Clip-On Current Sensor A DC and A (DC+AC)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(V input), all current displays appear with the correct value in
accordance with the selected transformation ratio. The only
prerequisite is that the current sensor is equipped with at least
one of the below listed transformation ratios, and that the ratio
has been previously selected in the following menu (Cl ip 0FF).
Current Clip Setup Menu
1nFo
... SET rAtE ... CL IP
0ff/1 / 10/100/1000

Maximum allowable operating voltage is equal to the current
transformers nominal voltage.
When reading the measured value,
additional error resulting from the clip-on current sensor must also be
taken into consideration.
(default value: Cl ip = 0ff = voltage display)
Transf. Ratio
CL IP
Measuring Ranges Clip Type
300 mV 3 V 30 V
1:1
1 mV / 1 mA
300.0 mA 3.000 A 30.00 A WZ12C
1:10
1m V / 10 mA
3.000 A 30.00 A 300.0 A
WZ12B,
Z201A/B
1:100
1m V / 100 mA
30.00 A 300.0 A 3,000 kA
Z202A/B
METRAFLEX
1:1000
1 mV/1 A
300.0 A 3,000 kA 30.00 kA
Z202A/B, Z203A/B,
WZ12C
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
~
R
x
Current
0035
.
0
A
1:1000
A
DC
0035
.
0
A
1:1000
A
DC
AC
TRMS
FUNC
ENTER
R
i
R
i
~ 9 M
V
V
MEASURE
SETUP
... SET ... CL IP
FUNC
ENTER
FUNC
ENTER
 0ff
Measuring Ranges:
mA / A: see table