User manual

GMC-I Gossen-Metrawatt GmbH 35
Measurements: V/Hz, Ω, , Temp (°C/°F), , MΩ
@UISO
, A/Hz
5.3.2 Measurement with Thermocouples, Temp TC
Set the rotary switch to “Temp
RTD
”.
Note
The last selected temperature measurement or the last
selected temperature sensor, i.e. type K or Pt100/Pt1000,
remains in memory and is accordingly displayed. Press the
FUNC | ENTER key in order to change to the other
measuring function if required.
The reference temperature is measured at the internal
reference junction (see parameter “1tEMP” in Seite 50 regarding
querying).
Note
The internal reference temperature (temperature of the
internal reference junction) is measured by a temperature
sensor inside of the instrument. This may be somewhat
above room temperature as a result of internal heat-up, or
moving from warmer to colder surroundings or vice versa.
Connect the sensor to the two accessible jacks. The
instrument displays the measured temperature using the
selected unit of measure.
COM mA A
V
Temp
Ω MΩISO
Temp
RTD
0023
.
3
°
C
Temp
TC
Type K
TC
TE (TC)
0256
.
0
°
C
Pt1000
RTD
Measuring Range
TE (TC) K (NiCr-Ni) 250.0 +1372.0° C
FUNC
ENTER