Operating Instructions SECULIFE⏐HIT METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO⏐BASE TRMS Digital Multimeter 3-349-352-03 11/7.
Standard Equipment Contact Persons Standard Equipment 1 multimeter 1 KS17-2 measurement cable set 2 batteries 1 abbreviated operating instructions 1 CD ROM (contents: amongst other topics operating instructions and data sheet) 1 DAkkS calibration certificate Function SECULIFE⏐HIT METRAHIT⏐ METRAHIT⏐ METRAHIT⏐ PRO BASE METRAHIT⏐ TECH X-TRA/ OUTDOOR — • — • — mV/A mA/A — mV/A Ri = 1 MΩ — mV/A mA/A — mV/A Ri = 1 MΩ — mV/A mA/A — mV/A Ri = 1 MΩ ... 30 kHz ... 30 kHz ... 30 kHz ...
Contact Persons Accessories (sensors, plug inserts, adapters, consumable materials) The accessories available for your instrument are checked for compliance with currently valid safety regulations at regular intervals, and are expanded as required for new applications.
Standard Equipment Recalibration Service We calibrate and recalibrate all instruments supplied by GMC-I Messtechnik GmbH, as well as other manufacturers, at our service center, for example after one year within the framework of your test equipment monitoring program, as well as prior to use etc. Repair and Replacement Parts Service Calibration Laboratory* and Rental Instrument Service If required please contact: GMC-I Service GmbH Service Center Thomas-Mann-Str.
Contact Persons GMC-I Messtechnik GmbH 5
Table of Contents Contents Page Contents Page 1 Safety Features and Precautions ....................................... 8 5 1.1 1.2 1.3 Use for Intended Purpose ................................................................10 Meanings of Danger Symbols ..........................................................10 Meanings of Acoustic Warning Signals ............................................10 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols . 12 3 Initial Start-Up ...........
Table of Contents Contents Page 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only) .............................. 42 5.7.3 Direct and Pulsating Current Measurement with Clip-On Current Sensor, A DC and A (DC+AC) (METRAHIT⏐TECH and METRAHIT⏐BASE only) .....43 5.7.4 Alternating Current Measurement with Clip-On Current Sensor, A AC and Hz (METRAHIT⏐TECH and METRAHIT⏐BASE only) ............ 44 5.7.
Safety Precautions 1 Safety Features and Precautions You have selected an instrument which provides you with a high level of safety. This instrument fulfills the requirements of applicable European and national EC directives. This is confirmed by means of the CE mark. A corresponding declaration of conformity can be requested from GMC-I Messtechnik GmbH.
Safety Precautions • Be prepared for the occurrence of unexpected voltages at devices under test (e.g. defective devices). For example, capacitors may be dangerously charged. • Make certain that the measurement cables are in flawless condition, e.g. no damage to insulation, no interruptions in cables or plugs etc. • No measurements may be made with this instrument in electrical circuits with corona discharge (high-voltage). • Special care is required when measurements are made in HF electrical circuits.
Safety Precautions 1.1 Use for Intended Purpose • The respective multimeter is a portable device which can be held in the hand during the performance of measurements. • Only those types of measurements described in chapter 5 may be performed with the measuring instrument. • The measuring instrument, including measurement cables and plug-on test probes, may only be utilized within the specified measuring category (see page 59 and the table on page 8 regarding significance).
Safety Precautions GMC-I Messtechnik GmbH 11
Operating Overview – Connections, Keys, Rotary Switch, Symbols 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols 16 → chapter 7 15 chapter 3.1← * 1 → chapter 4.3 chapter 6← 2 → chapter 4.1.2 15 chapter 3← 14 3 → chapter 3 12 4 → chapter 5 chapter 6← 5 → chapter 6 12 chapter 4.4← 6 → chapter 3 ff. 7 → chapter 1.2 8 → chapter 3 9* → chapter 4.1 10 chapter 4.1 ← Max.
Operating Overview – Connections, Keys, Rotary Switch, Symbols Symbols Used in the Digital Display 1 2 3 4 5 6 7 8 9 10 12 12 24 23 14 15 22 15 16 20 Battery level indicator 21 1 2 3 4 5 6 7 8 9 10 11 12 17 19 18 Battery full Battery OK 13 14 15 16 Battery weak Battery (almost) dead, U < 1.
Operating Overview – Connections, Keys, Rotary Switch, Symbols Symbols of Rotary Switch Positions Switch FUNC Display Measuring Function V~ Hz (V) V~ 1kHz Hz (V) 1kHz V~ 1 MΩ V~ 1kHz Hz (V) 1kHz Hz (V) 1 MΩ V V MHz % 0/4 1 2 3 0/4 1 2 3 0/2 1 0/2 1 — Alternating voltage, TRMS AC, full bandwidth Voltage frequency, full bandwidth Alternating voltage, TRMS AC, with low-pass (1 kHz) Voltage frequency, with low-pass (1 kHz) Alternating voltage, TRMS AC, full bandwidth, input 1 MΩ Alternating voltage, TRMS A
Operating Overview – Connections, Keys, Rotary Switch, Symbols User Interface Symbols in the Following Chapters ... Scroll through main menu ... Scroll through sub-menu Select decimal point Increase/decrease value time Sub-menu/parameter (7-segment font) 1nFo Main menu (7-segment font, boldface) s IR t 5 V / 600 mA The device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.
Initial Start-Up – Setup 3 Initial Start-Up 3.1 Inserting Batteries or Rechargeable Batteries Be certain to refer to chapter 9.2 regarding correct battery installation. Momentary battery voltage can be queried in the Info menu (see chapter 6.3). ! Attention! Disconnect the instrument from the measuring circuit before opening the battery compartment lid in order to replace the batteries. Operation with the Power Pack (accessory for METRAHIT⏐X-TRA, not included, see chapter 10.
Initial Start-Up – Setup 3.4 Switching the Instrument Off Switching the Instrument Off Manually Ð Press the ON / OFF | LIGHT key until 0FF appears at the display. ON / OFF LIGHT Shutdown is acknowledged with a brief acoustic signal. Automatic Shutdown The instrument is switched off automatically if the measured value remains unchanged for a long period of time (maximum measured value fluctuation of approx. 0.
Control Functions 4 Control Functions 4.1 Selecting Measuring Functions and Measuring Ranges The rotary switch is linked to the automatic socket blocking mechanism, which only allows access to two connector jacks for each function. Be certain to remove the appropriate plug from its respective jack before switching to and from the “A” functions.
Control Functions 4.1.3 Quick Measurements Measurements performed using a suitable fixed measuring range are executed more quickly than those which utilize automatic range selection. Quick measurement is made possible with the following two functions: • Manual measuring range selection, i.e. selection of the measuring range with the best resolution (see chapter 4.1.2). or • With the DATA function (see chapter 4.4).
Control Functions Ð Briefly press the ZERO | ESC key. The instrument acknowledges storage of the reference value with an acoustic signal, and the “ZERO ΔREL” or the “ΔREL” symbol appears at the LCD. The value measured at the moment the key is pressed serves as a reference value. Ð The reference value can be cleared by once again pressing the ZERO | ESC key. Notes Regarding Relative Measurement • Relative measurement effects the digital display only.
Control Functions 4.4 Measured Value Storage: DATA (auto-hold / compare) An individual measured value can be automatically “frozen” with the DATA function (auto-hold). This is useful, for example, when contacting the measuring points with the test probes requires your full attention.
Control Functions Example The voltage measuring range is set manually to 10 V. The first measured value is 5 V, which is stored to memory because it is greater than 10% of the measuring range (= 1 V), and is thus reliably above the background noise level. As soon as the measured values drops to less than 10% of the measuring range, i.e. amounts to less than 1 V which corresponds to removal of the test probes from the measuring point, the instrument is ready to store a new value. 4.4.
Control Functions 4.5 Measurement Data Recording (METRAHIT⏐X-TRA⏐OUTDOOR only) The METRAHIT⏐X-TRA⏐OUTDOOR is capable of recording measurement data using an adjustable sampling rate for long periods of time in the form of measurement series. Data are stored to a battery backed memory module, and are retained even after the multimeter is switched off. The system acquires measured values relative to real-time. Stored measured values can subsequently be read out with the help of METRAwin 10 software.
Control Functions During Recording STORE is displayed underneath the analog display during memory mode operation, and memory occupancy can be controlled: Querying Memory Occupancy Memory occupancy can be queried during recording with the help of the “1nFo ” menu (see also chapter 6.3). Memory occupancy range: 000.1% to 099.9%. StoP 000.3% MEASURE SETUP The following message appears as soon as memory is full: “100.0%”.
Control Functions GMC-I Messtechnik GmbH 25
V/Hz, Ω, Temperature, 5 Measurements 5.1 Voltage Measurement and A/Hz Measurements Notes Regarding Voltage Measurement • The multimeter may only be operated with installed batteries or rechargeable batteries. Dangerous voltages are otherwise not indicated, and the instrument may be damaged. • The multimeter may only be operated by persons who are capable of recognizing contact hazards and taking the appropriate safety precautions.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC) Note Only to observe for METRAHIT⏐TECH: Set the CL iP parameter to 0FF in the current clip setup menu. Otherwise all measured values are displayed in amperes, and are corrected by the amount resulting from the selected transformation ratio for an interconnected clip-on current sensor. MEASURE SETUP FUNC ENTER 1nFo ... SET FUNC ENTER 1 /10/100/1000 / 0ff V 020.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.1.2 Alternating Voltage Measurement with 1 MΩ Load Resistance and Frequency Measurement with Selectable Low-Pass Filter (METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only) The measuring instrument includes a V1MΩ switch position for electricians with an input resistance of approximately 1 MΩ. Erroneous displays resulting from capacitive coupling during voltage measurement in power supply systems are reduced to a minimum in this way.
V/Hz, Ω, Temperature, VΩ Temp Measuring Ranges: V~: 100 mV…1000 V Max. 1000 V (< 10 kHz) Max. 100 V (> 10 kHz) Hz: 1 Hz … 100 kHz Pmax = 3 x 106 V x Hz for U > 100 V COM ~ Warnings regarding dangerous voltage: > 55 V AC or > 70 V DC: V~ 1MΩ 1 kHz V~ with Filter ! 0225.6 V V 230.0 Hz with Filter ! AC TRMS V~ FUNC ENTER 0226.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter (METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only) Note METRAHIT⏐TECH: see note in chapter 5.1.1. Ð In accordance with the voltage or frequency to be measured, turn the rotary switch to V~ or Hz. Ð Connect the measurement cables as shown. The “⊥” connector jack should be grounded.
V/Hz, Ω, Temperature, VΩ Temp Measuring Ranges: V~: 100 mV…1000 V Max. 1000 V (< 10 kHz) Max. 100 V (> 10 kHz) Hz: 1 Hz … 100 kHz Pmax = 3 x 106 V x Hz for U > 100 V COM ~ V~ Hz V~ AC TRMS FUNC ENTER 050.03 AC Hz Hz 230.0 > 1000 V: Voltage Comparator for Displaying Dangerous Voltage The input signal or measuring signal is checked by a voltage comparator for dangerous spikes, because these do not appear at the display when the low-pass filter is used.
and A/Hz Measurements 5.1.5 Voltage Measurements at Above 1000 V Voltages of greater than 1000 V can be measured with a high-voltage probe, e.g. the HV31) or the HV302) from GMC-I Messtechnik GmbH. It is absolutely essential to earth the ground terminal in this case. Observe all applicable safety precautions! Voltage Measurements at Above 1000 V with the HV3 High-Voltage Probe VΩ Temp Red 5.1.
V/Hz, Ω, Temperature, 5.1.6 Frequency and Duty Cycle Measurements (METRAHIT⏐X-TRA⏐OUTDOOR only) Ð Set the rotary switch to MHz or %. and A/Hz Measurements MHz MHz 1.0323 Ð Connect the measurement cables as shown. Make sure that a current measuring range (“A”) has not been activated when the multimeter is connected for frequency or duty cycle measurement! fP MHz % FUNC ENTER U ! Attention! The applied signal voltage may not exceed 5 V. 002.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.2 Resistance Measurement, Ω Ð Disconnect supply power from the electrical circuit of the device to be measured, and discharge all high-voltage capacitors. Ω 000.00 Ð Make sure that the device under test is voltage-free. Interference voltages distort measurement results! Refer to chapter 5.1.1 regarding testing for the absence of voltage with the help of the direct voltage measurement. kΩ Measuring Ranges: 100 Ω … 40 MΩ Ð Set the rotary switch to “Ω”.
V/Hz, Ω, Temperature, 5.3 Continuity Test Ð Disconnect supply power from the electrical circuit of the device to be measured, and discharge all high-voltage capacitors. 000.8 Ð Make sure that the device under test is voltage-free. Interference voltages distort measurement results! Ð Set the rotary switch to Ω . FUNC ENTER Ð Connect the conductor path under test as shown.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.4 Diode Testing with a Constant Current of 1 mA Ð Disconnect supply power from the electrical circuit of the device to be measured, and discharge all high-voltage capacitors. 000.8 Ð Make sure that the device under test is voltage-free. Interference voltages distort measurement results! Refer to chapter 5.1.1 regarding testing for the absence of voltage with the help of the direct voltage measurement. Ð Set the rotary switch to Ω FUNC ENTER .
V/Hz, Ω, Temperature, 5.5 Temperature Measurement Temperature measurement is performed with a type K thermocouple (accessory, not included), which is connected to the voltage input. Alternatively, a Pt100 or Pt1000 resistance thermometer can be used with the METRAHIT⏐X-TRA⏐OUTDOOR . Selecting the Unit of Measure for Temperature MEASURE SETUP 1nFo ...
V/Hz, Ω, Temperature, and A/Hz Measurements 5.5.2 Measurement with Resistance Thermometers (METRAHIT⏐X-TRA⏐OUTDOOR only) Ð Set the rotary switch to “TempTC” or “TempRTD”. Ð Upon pressing the FUNC | ENTER key, the selected value is activated and the display is returned to the measuring function. Cable resistance remains in memory even after the instrument has been switched off. The last selected temperature measurement or sensor, i.e.
V/Hz, Ω, Temperature, Capacitance Measurement (METRAHIT⏐X-TRA⏐OUTDOOR and METRAHIT⏐TECH only) Ð Disconnect supply power from the electrical circuit of the device to be measured, and discharge all high-voltage capacitors. and A/Hz Measurements 5.6 Ð Make sure that the device under test is voltage-free. Capacitors must always be discharged before measurement is performed. Interference voltages distort measurement results! Refer to chapter 5.1.
V/Hz, Ω, Temperature, 5.7 and A/Hz Measurements Current Measurement Notes Regarding Current Measurement • The multimeter may only be operated with installed batteries or rechargeable batteries. Dangerous currents are otherwise not indicated, and the instrument may be damaged. • Set up the measuring circuit in a mechanically secure fashion, and secure it against inadvertent breaks. Select conductor cross-sections and lay out connections such that they do not overheat.
V/Hz, Ω, Temperature, 5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) (METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only) Ð First disconnect supply power from the measuring circuit or the power consumer (1), and discharge any capacitors. A A Current measurement may only be performed with installed batteries! Ð In accordance with the current to be measured, turn the rotary switch to A or A .
V/Hz, Ω, Temperature, and A/Hz Measurements 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only) Ð First disconnect supply power from the measuring circuit or the power consumer (1), and discharge any capacitors. Ð In accordance with the current or frequency to be measured, turn the rotary switch to A~ or Hz. Ð Select the desired measured quantity by briefly pressing the FUNC | ENTER multifunction key.
V/Hz, Ω, Temperature, 5.7.3 Direct and Pulsating Current Measurement with Clip-On Current Sensor, A DC and A (DC+AC) (METRAHIT⏐TECH and METRAHIT⏐BASE only) Transformer Output, Voltage/Current When a clip-on current sensor is connected to the multimeter (METRAHIT⏐TECH: V input, METRAHIT⏐BASE: V input), all current displays appear with the correct value in accordance with the selected transformation factor.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.7.4 Alternating Current Measurement with Clip-On Current Sensor, A AC and Hz (METRAHIT⏐TECH and METRAHIT⏐BASE only) (default setting: METRAHIT⏐TECH: OFF, METRAHIT⏐BASE: 1:1000) Transformer Output, Voltage/Current When a clip-on current sensor is connected to the multimeter (METRAHIT⏐TECH: V input, METRAHIT⏐BASE: V input), all current displays appear with the correct value in accordance with the selected transformation factor.
V/Hz, Ω, Temperature, and A/Hz Measurements 5.7.5 Direct, Pulsating and Current Measurement with Clip-On Current Transformer A DC, A (DC+AC), A AC and Hz (METRAHIT⏐TECH only) Transformer Output, Current/Current When a clip-on current transformer is connected to the multimeter (METRAHIT⏐TECH: mA/A input), all current displays appear with the correct value in accordance with the selected transformation factor.
Device and Measuring Parameters 6 Device and Measuring Parameters The instrument’s “SET ” mode (menu mode) makes it possible to set operating and measuring parameters, query information and activate the interface. Ð The menu mode is accessed by pressing the MEASURE | SETUP key, assuming that the instrument is switched on and set to “Measure” (measuring mode operation). “1nFo ” appears at the display. Ð The main menus, i.e.
Device and Measuring Parameters 6.1 Paths to the Various Parameters 020.00 V DC AC TRMS MEASURE SETUP 6.2 1nfo Main Menus → 1nFo StorE * SEnd * List of All Parameters Parameter XTRA TECH PRO BASE 0.diSP • • • • • — — — Addr A.diSP • • • • SET tEMP APoFF • • • • bAtt bEEP CLEAr • • • • 49: A.
Device and Measuring Parameters 6.3 Querying Parameters – InFo Menu (as moving letters) bAtt – query battery voltage FUNC MEASURE 1nFo ENTER bAtt: 2.75 V. SETUP vErSion – query firmware version FUNC MEASURE 1nFo ENTER bAtt: vErSion: 2.09 SETUP ItEMP – query reference temperature The temperature of the internal reference junction is measured with a temperature sensor in close proximity to the input sockets. FUNC MEASURE 1nFo ENTER bAtt: ...
Device and Measuring Parameters A.diSP – analog display: select display mode One of two different display modes can be selected for the analog display: • bArG: bar graph • Point: pointer MEASURE SETUP 1nFo ... SET FUNC ENTER bArG / Point FUNC ENTER rAtE ... A.diSP FUNC ENTER 10 ... 59 min on FUNC ENTER 1nFo ... SET FUNC ENTER 1, 10, 20, 30, 40, 90 Ω rAtE ...
Device and Measuring Parameters dAtE – enter date Entering the current date makes it possible to acquire measured values in real-time. MEASURE SETUP FUNC ENTER 1nFo ... SET FUNC ENTER FUNC ENTER FUNC ENTER Date and time must be reentered after replacing the batteries. tiME – set time Entering the correct time makes it possible to acquire measured values in real-time. MEASURE SETUP FUNC ENTER 1nFo ... SET FUNC ENTER 10:24 (hh:mm) See chapter 5.7.3 ff. rAtE ... dAtE 31.
Device and Measuring Parameters GMC-I Messtechnik GmbH 51
Interface Operation 7 Interface Operation (METRAHIT⏐X-TRA⏐OUTDOOR only) The METRAHIT⏐X-TRA⏐OUTDOOR is equipped with an infrared interface for the transmission of measurement data to a PC. Measured data are optically transferred through the instrument housing by means of infrared light to an interface adapter (accessory), which is attached to the multimeter. The adapter’s USB interface allows for the establishment of a connection to the PC via an interface cable.
Interface Operation 7.2 Configuring Interface Parameters irStb – status of the infrared receiver in the stand-by mode There are two possible switching statuses for the infrared interface when the multimeter is switched off: iron: IR appears at the display and the infrared interface is active, i.e. signals such as making commands can be received, and power is consumed even though the multimeter is switched off.
Technical Data 8 Technical Data Meas. Function Measuring Range V 100 mV 1 V 10 V 100 V 1000 V 11,999 TECH PRO X-TRA/OUTDOOR 100 μA A 1 mA X-TRA 10 mA 100 mA OUTDOOR 1 A PRO 10 A 10 mA A 100 mA 1 A TECH 10 A Factor: 1:1/10/100/1000 0.1/1/10/100 A A 1/10/100/1000 A TECH 10/100/1000/10000A 0.1/1/10/100 A A 1/10/100/1000 A TECH BASE 10/100/1000/10000A Ω 1) 2) 3) 54 Resolution at Upper Range Limit 100 1 10 100 1 10 40 100 5.
Technical Data Meas. Function F X-TRA OUTDOOR TECH Hz (V) Hz (A) Hz (A ) Hz (V) Hz (A) MHz X-TRA OUTDOOR % X-TRA OUTDOOR Measuring Range Resolution at Upper Range Limit 11,999 10 100 1 10 100 1000 nF nF μF μF μF μF 100.00 Hz 1.0000 kHz pF pF nF nF nF μF Discharge resistance 10 MΩ 1 MΩ 100 kΩ 12 kΩ 3 kΩ 3 kΩ 0.01 Hz 0.1 Hz Value U0 max 0.7 0.7 0.7 0.7 0.7 0.7 fmin 7) ±(... % rdg. + ... d) 1 + 6 6) with ZERO function active 1 + 6 6) 1 + 6 6) 1 + 6 6) 5 + 6 6) 5 + 6 6) ±(... % rdg. + ...
Technical Data Influencing Quantities and Influence Error Influencing Quantity Temperature 1) Sphere of Influence 0 °C ... +21 °C and +25 °C ... +40 °C Measured Quantity / Measuring Range 1) Influence Error (...% rdg. + ... d) / 10 K V 0.2 + 10 V 0.4 + 10 100 Ω ... 1 MΩ 0.5 + 10 > 1 MΩ 1 + 10 mA/A 0.5 + 10 mA/A 0.8 + 10 10 nF ... 100 μF 1+5 Hz 0.2 + 10 °C/°F (Pt100/Pt1000) 0.5 + 10 °C/°F thermocouple K 0.2 + 10 Influencing Quantity Measuring Quantity Influence Error (...% rdg.
Technical Data Influencing Quantity Crest factor CF 5) Sphere of Influence Measured Quantity / Measuring Range 1 ... 3 V > 3 ... 5 Influence Error 5) ± 1 % rdg. ,A ± 3 % rdg. Except for sinusoidal waveshape Influencing Qty. Sphere of Infl. Measured Quantity Influence Error V, A, Ω, Hz, °C 1 x intrinsic error Reference Conditions +23 °C ±2 K 40 ... 75% 45 ... 65 Hz sine 3 V ±0.1 V Ambient temperature Relative humidity Measured qty. frequency Measured qty.
Technical Data Data Interface (METRAHIT⏐X-TRA⏐OUTDOOR only) Power Supply Type Battery Optical via infrared light through the housing Data transmission (data transfer) Protocol Baud Rate Functions Serial, bidirectional (not IrDa compatible) Device specific 38,400 baud – Select/query measuring functions and parameters – Query/transmit momentary measurement data – Read out stored measurement data The USB X-TRA plug-in interface adapter (see accessories) is used for adaptation to the PC’s USB port.
Technical Data Display Acoustic Signals LCD panel (65 mm x 36 mm) with analog and digital display including unit of measure, type of current and various special functions For voltage For current Background illumination Background illumination is switched off approximately 1 minute after it has been activated. Fuse for METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO Analog Display Scaling Polarity display Overflow display Measuring rate Digital Display / Char.
Technical Data Electromagnetic Compatibility (EMC) Mechanical Design Interference emission EN 61326-1:2006, class B Interference immunity EN 61326-1:2006 EN 61326-2-1:2006 Housing Dimensions Weight Protection Ambient Conditions Operating temp. range TA Storage temp. range Relative humidity Elevation Deployment 60 −10 °C ... +50 °C −25 °C ... +70 °C (without batteries) Max.75%, no condensation allowed METRAHIT OUTDOOR only: max.
Technical Data GMC-I Messtechnik GmbH 61
Maintenance – Calibration 9 Maintenance and Calibration ! 9.1 Attention! Disconnect the instrument from the measuring circuit before opening the battery compartment lid or fuse cover in order to replace batteries or fuses! Displays – Error Messages Message Function Meaning FUSE Current measurement Blown fuse In all operating modes Battery voltage has fallen below 1.8 V Measurement Indicates overflow 0L 9.
Maintenance – Calibration Replacing the Batteries ! Attention! Disconnect the instrument from the measuring circuit before opening the battery compartment lid in order to replace the batteries. • In the active current measuring range when voltage is applied If the fuse is blown or has not been inserted, “FuSE” appears at the digital display. The fuse interrupts the current measuring ranges. All other measuring ranges remain functional. Ð Set the instrument face down onto the working surface.
Maintenance – Calibration ! Attention! Use specified fuses only! If fuses with other blowing characteristics, other current ratings or other breaking capacities are used, the operator is placed in danger, and protective diodes, resistors and other components may be damaged. The use of repaired fuses or short-circuiting the fuse holder is prohibited.
Maintenance – Calibration 9.6 Recalibration The respective measuring task and the stress to which your measuring instrument is subjected affect the ageing of the components and may result in deviations from the guaranteed accuracy. If high measuring accuracy is required and the instrument is frequently used in field applications, combined with transport stress and great temperature fluctuations, we recommend a relatively short calibration interval of 1 year.
Accessories 10 Accessories 10.1 General The extensive accessories available for our measuring instruments are checked for compliance with currently valid safety regulations at regular intervals, and are expanded as required for new applications.
Accessories 10.4 Interface Accessories for METRAHIT⏐X-TRA⏐OUTDOOR (not included) USB X-TRA Bidirectional Interface Adapter This adapter makes it possible to connect STARLINE multimeters which are equipped with a serial IR interface to the USB port at a PC. The adapter allows for data transmission between the multimeter and the PC.
Index 11 Index Numerics D 0.diSP ................................................................... 48 dAtE ................................................................. 48, 50 Default Settings .................................................... 50 Diode Test ............................................................. 36 Display Illumination .............................................. 16 Duty Cycle Measurement ..................................... 33 Automatic ..............................
Index Standard Equipment .............................................. 2 Switching the Instrument On Manual ..........................................................16 Via PC ............................................................16 Symbols Digital Display ...............................................13 Instrument .....................................................15 Rotary Switch Positions ................................14 T Temperature Measurement With Resistance Thermometers ................
Edited in Germany • Subject to change without notice • A pdf version is available on the internet GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg • Germany Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.