Operation Manual

GMC-I Messtechnik GmbH 35
V/Hz, Ω, Temperature, and A/Hz Measurements
5.3 Continuity Test
Ð Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
Ð Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Ð Set the rotary switch to .
Ð Connect the conductor path under test as shown.
Depending upon the selected limit value, the multimeter
generates a continuous acoustic signal in the case of continuity or
short-circuit, i.e. at a value of less than the selected limit value.
0L” appears at the display in the case of an open connection.
The limit value can be adjusted in the “SETUP” menu (see also
chapter 6.4):
1nFo
... SET rAtE ... bEEP
1, 10, 20, 30, 40, 90 Ω

(10 = default setting)
MEASURE
SETUP
FUNC
ENTER